19
Depending on the type of current sensor, MiniFlex
®
/AmpFlex
®
, MN clamp, or adapter unit, enter the nominal primary current. To
do this, press the
Enter
key. Use the
,
,
and
keys to choose the current.
AmpFlex
®
A196A or A193 and MiniFlex
®
MA193, MA194 or MA196: 100, 400, 2,000 or 10,000A (depending on the sensor)
PAC93 clamp and C193 clamp: automatic at 1,000A
MN93A clamp, 5A range, 5A Adapter: 5 to 25,000A
MN93A clamp, 100A range: automatic at 100A
MN93 clamp: automatic at 200A
E3N clamp: 10 or 100A
J93 clamp: automatic at 3,500 A
Validate the value by pressing the
Enter
key.
3.5.6. NOMINAL PRIMARY CURRENT OF THE NEUTRAL
Press the
key to go to the next screen.
If you connect a current sensor to the current terminal of the neutral, enter its nominal primary current too in the same way as before.
3.5.7. AGGREGATION PERIOD
Press the
key to go to the next screen.
Figure 16
To change the aggregation period, press the
Enter
key, then use the
and
keys to choose the value (1 to 6, 10, 12,
15, 20, 30, or 60 minutes).
Validate by pressing the
Enter
key.
Summary of Contents for PEL 106
Page 1: ...Power energy logger EN User s manual PEL 106 ...
Page 34: ...34 Three phase 3 wire balanced 3P 3W b U12 U23 U31 f P Q S PF P Q S tan ϕ ϕ I1 U12 I1 I2 I3 ...
Page 36: ...36 Three phase 4 wire wye balanced 3P 4WYb I1 I2 I3 V1 V2 V3 VN U12 U23 U31 f P Q S tan ϕ ...
Page 37: ...37 DC 2 wire dC 2W DC 3 wire dC 3W P Q S PF ϕ I1 V1 P Q S tan ϕ P I V VN I1 I2 IN ...
Page 38: ...38 DC 4 wire dC 4W P I1 I2 I3 IN V1 V2 V3 VN V1 V2 VN ...
Page 46: ...46 Two phase 3 wire 1P 3W I1 I2 V1 V2 U12 VN P Q S P Q S ...
Page 49: ...49 P Q S ...