58
b) MiniFlex
®
MA193, MA194 and MA196
MiniFlex
®
MA193 and MA196
Nominal range
100 / 400 / 2,000 A
ac
Measurement range
200 mA to 2,400 A
ac
Maximum clamping diameter
Length = 250 mm; Ø = 70 mm (MA 193 only)
Length = 350 mm; Ø = 100 mm
Influence of the position of the
conductor in the sensor
≤
1.5% typical, 2.5% maximum
Influence of an adjacent conductor
carrying an AC current
> 40 dB typical at 50/60 Hz for a conductor touching the sensor and
> 33 dB near the snap
Safety
IEC 61010-2-032, degree of pollution 2, 600V CAT IV, 1,000V CAT III
Table 11
Remark:
Currents < 0.05 % of the nominal range will be set to zero.
The
nominal ranges are reduced to 50/200/1,000/5,000 A
ac
at 400 Hz.
MiniFlex
®
MA194
Nominal range
100 / 400 / 2 000 / 10 000A
ac
(for the 1000 mm model)
Measurement range
50 mA to 2,400 A
ac
Maximum clamping diameter
Length = 250 mm; Ø = 70 mm
Length = 350 mm; Ø = 100 mm
Length = 1 000 mm, Ø = 320 mm
Influence of the position of the
conductor in the sensor
≤
2,5 %
Influence of an adjacent conductor
carrying an AC current
> 40 dB typical at 50/60 Hz for a conductor touching the sensor and
> 33 dB near the snap
Safety
IEC 61010-2-032, degree of pollution 2, 600V CAT IV, 1,000V CAT III
Table 12
Remark:
Currents < 0.05 % of the nominal range will be set to zero.
The
nominal ranges are reduced to 50/200/1,000/5,000 A
ac
at 400Hz.
The 10,000A range operates provided that the conductor can be clamped in the MiniFlex
®
sensor.
c) PAC93 clamp
Remark:
The power calculations are set to zero while the current zero is adjusted.
PAC93 clamp
Nominal range
1,000 A
ac
, 1,300 A
dc
93
Measurement range
1 to 1,000 A
ac
, 1 to 1,300 A
peak ac+dc
Maximum clamping diameter
One 42mm conductor or two 25.4mm conductors, or two 50 x
5mm bus bars
Influence of the position of the con
-
ductor in the clamp
> 40 dB typical at 50/60 Hz
Influence of an adjacent conductor
carrying an AC current
> 40 dB at 50/60Hz
Safety
IEC 61010-2-032, degree of pollution 2, 300 V CAT IV, 600 V CAT III
Table 13
Remark:
Currents < 1 A
ac
/
dc
will be set to zero in AC networks.
Summary of Contents for PEL 106
Page 1: ...Power energy logger EN User s manual PEL 106 ...
Page 34: ...34 Three phase 3 wire balanced 3P 3W b U12 U23 U31 f P Q S PF P Q S tan ϕ ϕ I1 U12 I1 I2 I3 ...
Page 36: ...36 Three phase 4 wire wye balanced 3P 4WYb I1 I2 I3 V1 V2 V3 VN U12 U23 U31 f P Q S tan ϕ ...
Page 37: ...37 DC 2 wire dC 2W DC 3 wire dC 3W P Q S PF ϕ I1 V1 P Q S tan ϕ P I V VN I1 I2 IN ...
Page 38: ...38 DC 4 wire dC 4W P I1 I2 I3 IN V1 V2 V3 VN V1 V2 VN ...
Page 46: ...46 Two phase 3 wire 1P 3W I1 I2 V1 V2 U12 VN P Q S P Q S ...
Page 49: ...49 P Q S ...