background image

32

32

3.3 Audio continuity test (

 )

1. Connect the measurement cables to the unit terminals.

2. Set the rotary switch to position " 

 ".

3. Connect the unit to the circuit to be tested. The buzzer is

continuously active as soon as contact is established (circuit
closed) and if the resistance value measured is lower than
the threshold value chosen by the programming (adjustable
from 1 to 40 

, see § 03.12.10).

Above 400 

, the display indicates 

OL.

Cable resistance compensation (

 zero)

To  measure low resistance values, measure the cable
resistance first.
- Short-circuit the cables.
Press and hold the HOLD key until zero appears on the

display. The cable resistance value will then be memorised
and subtracted from the value of the resistance measured
later.

Note: If the value measured is higher than 2 

, this

correction is stopped and the memorised correction
value is reset to zero.

Characteristics

Display range

400 

Measuring range

0.0 to 399.9 

Accuracy

 (1):

1% R +2 pt

Resolution

0.1 

Open circuit voltage

 3.2 V

Measuring current

320 µA

Protection

500 V AC or 750 V (DC or peak)

(1) with compensation for measurement cable resistance

MIN/ MAX Mode:
- Accuracy: same as previous table +0.2% R
- Capture time: 100 ms typ.

3.4 Resistance measurements (

)

1. Connect the measurement cables to the unit terminals.

2. Set the rotary switch to position " 

 " and press once

on the yellow key: The 

 symbol is no longer displayed.

3. Connect the unit to the circuit to be tested.

Range selection is automatic.
To measure low resistance with accuracy, compensate the
cable measurement resistance 

(see § 3.3).

Above 40 

, the display indicates 

OL.

Summary of Contents for F 05

Page 1: ...Vielfachmesszange Pinza multimetro Pinza multim trica F 05 F R A N A I S E N G L I S H D E U T S C H I T A L I A N O E S P A N O L Notice de fonctionnement User s manual Bedienungsanleitung Libretto d...

Page 2: ...ecautions detailed PRECAUTIONS FOR USE Never use on networks with voltage higher than 600 V in relation to the ground and whose overvoltage category is higher than III i e fixed industrial and domesti...

Page 3: ...ode 39 3 12 6 Selections possible in continuity function 39 3 12 7 Selection OF INRUSH function 39 3 12 8 3 12 8 Suppression of automatic stop 39 3 12 9 Activation of the V Live function 40 3 12 10 Mo...

Page 4: ...cable resistance zero Compliance with CEI electrical safety standards and CE markings Lightness and ruggedness for field use With all new additional features Inrush function in particular for measurem...

Page 5: ...conductor test continuity test Power factor calculation Phase order indication measurement see description 3 11 MIN MAX operates by end around shift on short pressure MIN MAX V and A Functions Other f...

Page 6: ...display includes the digital display of the measured values the related units and symbols Digital display 4 digits 9999 points 3 decimal points and signs DC and peak measurement OL Positive value rang...

Page 7: ...high pitched sound prohibited key Short and low sound Quit MIN MAX mode 2 short and high pitched beeps validation of a configuration parameter Short and medium sound every 400 ms voltage measured hig...

Page 8: ...io indication is activated if the V Live function is selected see 3 12 9 Display range 40 V 400 V 4000 V 1 Measuring range 2 0 2 V to 40 0 V to 400 to 600 V 39 99 V 399 9 V 400 to 900Vpeak Accuracy 1...

Page 9: ...rom the value of the resistance measured later Note If the value measured is higher than 2 this correction is stopped and the memorised correction value is reset to zero Characteristics Display range...

Page 10: ...on to be tested The measurement current moves from the terminals to the COM terminal It corresponds to the direct testing of the semi conductor junction Short circuit junction audio indication for a t...

Page 11: ...1 A 1 In DC the display indicates OLabove 400 A and OL above 400 A 600 A in PEAK mode In AC the display indicates OL over 400 Arms 900 V in PEAK mode 2 In AC if the value of the current measured is 0...

Page 12: ...frequency 200 ms at 50 Hz Range for use 5 A peak for the first period of the signal 3 8 Power measurement W 1 Connect the measurement leads to the instrument s terminals complying with the polarities...

Page 13: ...the clamp configured in power measurement switch on position W and correctly connected see 3 8 perform a short press on the yellow key the power factor is displayed The power factor is by definition a...

Page 14: ...ion of a reference period on one phase L1 L2 for example 2 Integration of a reference period on one phase L1 L3 3 Calculation of the time delay between reference and measurement periods enabling deter...

Page 15: ...seconds after messages may be displayed leaving L2 after 10 s max MEAS determination of period for measurement in progress contact on L3 a was performed too late more than 10 s after display rEF OK E...

Page 16: ...key are used to select manually AC measurement DC measurement and return to automatic mode When the mode is selected manually the AC or DC symbol is fixed The selection of this manual mode is impossib...

Page 17: ...Hold the Yellow key pressed and bring the rotary switch from the OFF position to A position The unit emits a double beep then all the segments of the digital display and the symbol flash The default...

Page 18: ...ivation by switching from the OFF position of the switch or by actuating any key 4 4 Environmental parameters Temperature Humidity 0 10 20 30 40 50 60 70 80 90 50 40 30 20 10 0 10 20 30 40 50 60 70 80...

Page 19: ...as per IEC 68 2 6 Auto power OFF per UL94 Housing V0 Jaws V0 Display window V2 4 6 Variations in operating range Quantity Meas range Quantity Influence quantities quantities influenced Typical Maxi Ba...

Page 20: ...1 ct PF 0 7 and I 5 A 1 R 1 ct inductive and 0 5 and I 10 A W 0 5 R 3 R 1 ct capacitive 0 2 and I 20 A 8 R 1 ct Rejection of serial mode 0 600 V 50 Hz VDC 50 dB 45 dB in DC 0 400 A 50 Hz ADC 40 dB 35...

Page 21: ...00 6 00 8 00 100 Hz 1000 Hz 10000 Hz Specified limits Frequency Hz Error measurement I f f 10 Hz 25 00 20 00 15 00 10 00 5 00 0 00 5 00 10 00 100 Hz 1000 Hz 10000 Hz Specified limits Frequency Hz Err...

Page 22: ...0 88 set of 2 alligator clips NF EN 61010 P01 1018 48 set of 2 cables with IP2X touch prods P01 2951 57 carrying case N 7 P01 2985 32 Different measurement accessories widen the scope or provide new f...

Page 23: ...er 7 2 Storage If the clamp is not used for a period of more than 60 days remove the battery and store it separately 7 3 Cleaning The clamp must be disconnected from any external electrical source and...

Reviews: