–12 –
5.4 Using the alarm function
1. In the Material Setup menu, make sure that the MIN and MAX limit values have
been set for each material characteristic.
2. The limit value alarm has to be enabled in the main menu.
5.5 Cal. Adjustment
By performing a calibration adjustment, you can adjust a material characteristic
calibrated for a particular material to a different material or diameter without creating a
new material characteristi
Requirements:
1. Prepare the measuring setup as shown in figure below.
2. Select the desired material characteristic (section 5.2).
3. Move the tension meter into the desired measuring position and carry out a zero
adjustment, if required.
Adjusting:
1. Select the Cal. Adjustment menu item from the main menu (see figure below)
2. Insert the material to be measured (section 5.2) and push the thumbpiece into
the measuring position.
3. Press the or key to perform the calibration adjustment until the value on the
display corresponds to the weight suspended to the material. The adjustment can be
performed in 1% steps within the range from +10% to –1%.
4. Press the
button to save the determined value.
NOTE:
Please note that this is a general value. So determine it separately for each
material characteristic and note it down for later reference. It is not possible to save it
per material characteristic. To disable calibration adjustment, reset the value to 0% in the
Cal. Adjustment menu.
The calibration adjustment must be conducted with the weight unit that is
adjusted in the menu tension unit and carried out with weights corresponding
to the choosen unit.