–17 –
4) Use the
and
keys to scroll to the known thickness value.
NOTE:
The longer the keys are pressed and held, the faster the value will increment/
decrement.
NOTE:
Pressing the
key prior to pressing the
key will abort the CAL
routine without
saving any changes.
5. Once the known thickness value is being displayed, press the key to display the
calculated material velocity edit screen.
NOTE:
The calculated velocity can be edited, if needed, by pressing
the
and
keys to scroll and edit the velocity value.
6. Press the
key to set the calculated material velocity and return to the
measurement screen.
CHECK YOUR CALIBRATION! Place the transducer back on the calibration point
and verify the thickness. If the thickness is not correct, repeat the steps above.
Two Known Thicknesses
The two point calibration should be considered when an application requires improved
accuracy over a small measurement range based on tolerance requirements. This
calibration option calculates both the ‘probe zero’ and ‘velocity value. If the two point
option is used, a probe zero is not required. For example, if the measurement range was
.080” (2.03mm) to .250” (6.35mm), two known samples or locations on the test material
would be needed for the minimum and maximum boundaries of the test range. Using the
range above, a one point calibration would be performed at .250” (6.35mm) and a two
point calibration at .080” (2.03mm), or something close to the min/max values of the
measurement range.
NOTE:
The TI-25MX also offers the capability of setting the ‘probe zero’ to use any
reference standard as the ‘probe zero’ standard. For clarification, if it’s desired to use a
one inch reference of a specific material type as the ‘zero’ reference, performing the first
point of a two-point calibration sets the internal zero of the TI-25MX. This should be
used only in manual probe zero mode “on block”.
The following steps outline this procedure:
Two Point Calibration
1. Physically measure a minimum and maximum calibration point of the exact sample
material, or locations directly on the material to be measured, using a set of calipers
or a digital micrometer.
Summary of Contents for TI-25MX