– 3 –
2.0 KEYPAD, MENU, DISPLAY & CONNECTORS
The Keypad
2.1 ON/OFF/ENTER Key
The ON/OFF/ENTER key powers the unit ON or OFF. Since the same key is
also used as an ENTER key, the gauge is powered off by pressing and holding
down the key until the unit powers off.
Once the gauge is initially powered on, this key will function as the ENTER key, similar
to a computer keyboard. This key will be used to select or set a menu option.
NOTE:
Unit will automatically power off when idle for 5 minutes. All current settings
are automatically saved prior to powering off.
2.2 PRB 0 Key
The PRB 0 key is used to “zero” the TI-25MX in much the same way that a
mechanical micrometer is zeroed. If the gauge is not zeroed correctly, all of the
measurements that the gauge makes may be in error by some fixed value. Refer
to section 5.1 for a further explanation of this important feature.
2.3 CAL Key
The CAL key is used to enter and exit the TI-25MX’s calibration mode. This
mode is used to adjust the sound velocity value that the TI-25MX will use when
calculating thickness. The tool will either calculate the sound-velocity from a
sample of the material being measured, or allow a known velocity value to be entered
directly. This provides increased linearity between transducers. Refer to section 5.2 for
an explanation on the various calibration options.
Summary of Contents for TI-25MX
Page 1: ...MODEL TI 25MX ULTRASONIC THICKNESS GAUGE Operating Manual ...
Page 30: ......