–15 –
The procedure to enable/disable the lock feature is outlined below:
1. With the TI-25PX powered off, press and hold down the
key while
powering the TI-25PX on .
The lock icon will be illuminated on the
display.
2. To unlock the TI-25PX repeat step one, but hold down the
key while
powering the TI-25PX on.
6.8 Factory Defaults
The TI-25PX can be reset to factory defaults at any time to restore the original gauge
settings. This should only be used if the gauge is not functioning properly, or perhaps
multiple features have been enabled and a clean start is needed.
The procedure to reset the gauge is outlined below:
1. With the TI-25PX powered off, press and hold down the
and
keys
while powering the TI-25PX on.
NOTE:
Once the measurement screen is displayed the keys can be released.