background image

5.1 Preparation and Start-up

Before starting up the appliance, remove the protective wrapping.

Then carefully clean the working surface and the external parts with
lukewarm water and detergent, using a damp rag to remove all traces
of anti-rust material applied in the factory, then dry with a clean cloth. 

5.1.1 Start-up

Before using the appliance for the first time, thoroughly clean out
the tank (see the chapter 6.3 “Cleaning and taking care of the
machine”).

Before starting up the appliance, check that its specifications (cate-
gory and type of gas used) match those of the family and group of
the gas available locally. 

If not, it is necessary to adapt the appliance to the gas family or
group required (see paragraph 5.1.10 “Conversion and adaptation”). 

To start up the appliance, see the instructions for regular use.

5.1.2 Check of power

The appliances must be used with the specific injectors for the nomi-
nal power.  

The power may be:  

•  the nominal power indicated on the data plate of the appliance;

•  the reduced capacity power.  

These injectors are shown in table 1.

Nominal power is also obtained in respect of the supply pressure:

•  from 15 to 22.5 mbar for gases of the second family (G20/methane)

•  from 25 to 45 mbar for gases of the 3rd family  

(G30/butane, G31/propane) 

The appliance shall not be operated outside the above-mentioned
pressure ranges. 

To adjust power with reduced capacity, use the data in table 1. 

If you wish to further check the nominal power, you may do so by

using a gas meter according to the so-called "volumetric method".  

A simple inspection is usually enough to check if injectors are func-
tioning correctly. 

5.1.3 Checking the input pressure  

Input pressure should be measured using a fluid measuring gauge
(e.g. a gooseneck pipe, min. resolution 0.1 mbar).

Remove lock screw (pos. 22 fig. 1) from the pressure intake tube and
connect the gauge hose: once measurement is complete, replace the
screw and do a seal check using a leak detector spray.

5.1.4 Power check with volumetric method

Using a gas meter and a stopwatch, you can read the volume of gas
output per time unit. The correct volume corresponds to the value
"E" expressed in litres/hour (l/h) or litre/minute (l/min).  

The following formula is used to calculate the value of “E”: 

It is important measure the power when the appliance is in standby
status.  

The calorific power value can be requested from the local gas company. 

The nominal power and the minimum power with respect to the
nominal pressure are obtained by consulting the table for the
adjustment of the gas passage (table 1).

WARNING

There is no pre-adjustment device for the nominal
power.

E

 =

Power

Operating

 calorific value

5 - SET-UP FOR OPERATION

0820_GB_99

- GAS PASTA COOKER

7

· 12

01/2016

4

22

6

1

5

3

2

21

24

0

Fig. 1

Summary of Contents for CT74GPC

Page 1: ...GAS PASTA COOKER Instructions for installation use e maintenance 01 2016 CT74GPC...

Page 2: ...CHARACTERISTICS 0820_GB_99 GAS PASTA COOKER 2 12 01 2016 Supplied by Date Customer Service FAX e mail...

Page 3: ...tment 8 5 1 11 Replacement of burner injector 8 5 1 12 Replacement of burner pilot injector 8 5 1 13 Setting reduced capacity power 9 5 2 Maintenance 9 5 3 Replacing parts 9 5 3 1 Gas cock 9 5 3 2 The...

Page 4: ...DIAGRAM 1 4 4 3 1 7 1 67 6 4 4 16 4 3 4 4 D Gas 1 2 Water load pipe 12 Water drainpipe 1 GC female Gas 1 2 Water load pipe 12 Data plate Water drainpipe 1 GC female Gas 1 2 Water load pipe 12 Data pl...

Page 5: ...purposes Installation repair and use must be carried out by expert personnel These instructions for installation are for our gas pasta cooker set up for the category in the table 1 on pag 5 The data p...

Page 6: ...itting tiles or thermal radiation protection material to the walls Before connecting the appliance to the gas supply check on the data plate that the appliance is suitable and type tested for the type...

Page 7: ...pliance shall not be operated outside the above mentioned pressure ranges To adjust power with reduced capacity use the data in table 1 If you wish to further check the nominal power you may do so by...

Page 8: ...thus requiring a second operation check 5 1 10 Conversion and adjustment To change over form one kind of gas to another for example from methane to liquid gas or to another type of gas the use of suit...

Page 9: ...37 fig 3 To make the operation easier as well as all those involving the pilot burner the screws can be removed pos 33 fig 3 and the pilot burner brought to a more comfortable position 5 3 4 Burner Re...

Page 10: ...ht the pilot burner place the knob in the maximum or minimum position so that the main burner igni tes Burner power regulation must be done between the max and min supply position of easily selectable...

Page 11: ...amber only with food having a high salt content It is preferable to cook different food together i e fatty dishes or vegetables containing acids Avoid damaging the stainless steel surface in particula...

Page 12: ...THIS BOOKLET DUE TO COPYING OR PRINTING ERRORS DUE TO ITS POLICY OF CONTINUAL PRODUCT IMPROVEMENT THE MANUFACTURER RESERVES THE RIGHT TO MAKE ANY CHANGES DEEMED NECESSARY THE MANUFACTURER CANNOT BE HE...

Reviews: