Beads Low
Beads Medium
Beads High
** = The high Coefficient of Variation is primarely due to Poisson distribution.
Page 1 of 1
0
#####
%
Viability
Count
P/N: 071-0030
##### %
Count
This lot was released by:
##### x 10
5
beads/mL
x 10
5
beads/mL
Viability
#####
##### - #####
#####
x 10
6
beads/mL
91 - 97 %
-
#####
0 - 5 %
Application:
NC-202
PQ Kit
is part of the NucleoCounter® NC-202™ system and is used for performing
performance qualification (PQ) using the
Via2-Cassette™ and the NC-View® Software on NC-202™
.
For further information please see package insert.
- #####
Accepted Range
Variation Limits
x 10
4
beads/mL
#####
Results on Master Instrument
Count
x 10
4
beads/mL
P/N: 071-0029
#####
Viability
Certificate of Analysis for lot.
0
0
Jan 1900
Mmm-yyyy
912-2022
Release date:
Exp. Date:
NC-202
PQ Kit
Using the Via2-Cassette™ and Count & Viability Assay on the NucleoCounter®
NC-202™
Part number:
Lot number:
ChemoMetec A/S ∙ Gydevang 43 ∙ DK-3450 Allerod ∙ Denmark ∙ Support@chemometec.com ∙ www.chemometec.com
Doc. No.: 992-2024 v. 1
CV% < 28%**
SD < 2*
CV% < 11%**
CV% < 3%
CV% < 6%
CV% < 3%
* = When viability is close to 0, even insignificant changes in dead counts can result in extremely high CV%.
Instead, Standard Deviation (SD) is used to evaluate the result.
This lot has been tested according to the internal ChemoMetec quality procedures. The lot meets or exceeds all
product specifications.
##### x 10
6
beads/mL
95 - 100 %
%
P/N: 071-0031
EXAMPLE
Page 44 of 127