EN 12470-5

Clinical thermometers – Part 5: Performance of infra-red ear thermometers 

(with maximum device)

EN 62304

Medical device software - Software life-cycle processes

EN ISO 10993-1

Biological evaluation of medical devices - Part 1: Evaluation and testing within 

a risk management process

ASTM E 1965

Standard Speci

fi

cation for Infrared Thermometer for Intermittent 

Determination of Patient Temperature

16.  DISPOSAL

 

Dispose of the device in accordance with the regulation applicable at the place of 

operation. Dispose of at public collection point in the EU countries – 2002/96/EC WEEE 

Directive.

If you have any queries, please refer to the local authorities responsible for waste 

disposal.

NOTES:

t

  Handing of battery and wastes method, please act according to the native law to proceed to handle.

t

  Take out the battery if you are not going to use the unit for a long time.

To protect the environment, dispose of empty battery at your retail store or at appropriate 

collection sites according to national or local regulations.

Dispose of at public collection point in the EU countries – 2006/66/EC Directive.

17. NORMALIZED SYMBOLS

Attention: see Instructions for use!

Applied part of type BF

The batteries and electronic instruments must be disposed of in accordance 

with the locally applicable regulation, not with domestic waste.

  

Complies with the European Medical Device Directive (93/42/EEC), Noti

fi

ed 

Body is SGS United Kingdom Ltd.

    

Manufacturer information: The manufacturer Famidoc Technology Co., Ltd.

Authorized representative in the European Community. 

IP22

IP code of the device: this device’s grade of against ingress of solid foreign 

objects -- 

12.5mm diameter (and the against access to hazardous parts with 

fi

nger); the grade of waterproof is dripping (15° tilted).

 

Batch code

18.  Electromagnetic Compatibility (EMC) Tables

Guidance and manufacturer’s declaration - electromagnetic emissions

The customer or the user of the 4in1 should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment  - guidance

RF emissions 

CISPR11

Group 1

The device uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to 

causes any interference in nearby electronic equipment.

RF emissions 

CISPR11

Class B

The 4in1 device is suitable for use in all establishments 

other than domestic and those directly connected 

to the public low-voltage power supply network that 

supplies buildings used for domestic purposes.

Harmonic 

emissions 

lEC61000-3-2

Not applicable

Voltage 

lEC61000-3-3

Not applicable

Guidance and manufacturer’s declaration — electromagnetic immunity

The 4in1 device is intended for use in the electrom

The customer or the user of the 4in1 should assure that it is used in such an environment.

Immunity test

IEC 60601 Test 

level

Compliance 

level

Electromagnetic environment - 

guidance

Electrostatic 

discharge(ESD) 

lEC61000-4-2

±6kV contact

±8kV air 

 

±6kV contact

±8kV air 

Floors should be wood , concrete 

with synthetic material, the relative 

humidity should be at least 30%.

Guidance and - manufacturer’s declaration. Electromagnetic immunity

The 4in1

ow. 

The customer or the user of the 4in1 V1 should assure that it is used in such an environment. 

Immunity test

IEC 60501 Test 

level

Compliance lev el Electromagnetic environment - 

guidance 

C o n d u c t e d R F 

l E C 6 1 0 0 0 - 4 - 6 

Radiated  RF  lEC 

61000-4-3

3V rms

150 kHzto 

80 MHz  3V rms

3V/m  3V/m

80 MHz to

2.5 GHz

           

±6kV contact

±8k Vair 

Portable and mobile RF communications 

equipment should be used no closer 

to any part of the FDIR-V1 device,  

than the recommended separation 

distance calculated from the equation 

applicable to the frequency of the 

transmitter. Recommended separation 

distance.

 

 80 MHz to 800 MHz

 

800MHz to 2,5MHz

where P is the maximum output 

power rating of the transmitter 

in watts (W) according to the. 

transmitter manufacturer and d is the 

recommended separation distance in 

RF transmitters, as determined by an 

electromagnetic site survey, 

a. should be less than the compliance 

level in each frequency range. 

b. Interference may occur in the 

vicinity of equipment marked with the 
following symbol:   

NOTEI At 80 MHz end 800 MHz. the higher frequency range applies. 

a. Field strengths from 

fi

xed transmitters, such as base stations for radio(cellular/cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 

broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 

environment due to 

fi

xed RF transmitters, an electromagnetic site survey should be 

exceeds the applicable RF compliance level above, the should be observed to verify normal 

operation. If abnormal performance is observed, additional measures may be necessary, 

such as reorienting or relocating the FDIR-V1.

b. 

Recommended separation distances between portable and mobile RF communications 

equipment and the FDIR-V1 device.

FDIR-V1 device is intended for use in an electromagnetic environment in which radiated RF 

disturbances are controlled. The customer or the user of the FDIR-V1 device can help prevent 

electromagnetic interference by maintaining a minimum distance between portable and 

mobile RF communications equipment(transmitters) and the FDIR-V1 as recommended below, 

according to the maximum output power of the communication sequipment.

Rated maximum 

outpu

t power of 

transmitter W

Separation distance according to frequency of  transmitter m

Electrostatic 

discharge(ESD) 

lEC61000-4-2

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2,5 GHz

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

7.8

7.8

3.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, recommended separation 

distance d in meters(m) can be estimated using the equation applicable to the frequency of 

the transmitter, where P is the maximum output power rating of the transmitter in watts(W) 

accordable to the transmitter manufacturer.

NOTEI At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.

19. Software Version

This thermometer’s software version is Ver. 2.0.

20.  WARRANTY

1.  The warranty period for device is one year from date of delivery. In case of a warranty claim, the 

date of delivery has to be proven by means of the sales receipt or invoice.

2.  Repairs under warranty do not extend the warranty period.

3.  The following cases are excluded under the warranty

t

 All damage which has arisen due to improper treatment, e.g. nonobservance of the user 

instruction.

t

 

 

All damage which is due to repairs or tampering by the customer or unauthorized third parities.

t

 

 

Damage which has arisen during transport from the manufacturer to the consumer or during 

transport to the service centre.

t

 

 

Accessories which are subject to normal wear and tear.

4.  Liability for direct or indirect consequential losses caused by the unit is excluded even if the 

damage to the unit is accepted as a warranty claim.

 

The 

thermometer 

show the 

symbol “Hi” or 

“Lo” 

Temperature hampered by an air 

fl

ux.

Please leave the status and wait for 

30 minutes to measure. Re-measure 

according to the manual.

In the forehead measurement mode:

--Temperature readings too close together.

-- Measured the other object, such as the 

sunlight, the air from the 

fi

replace.

Hi: Higher than 42.9°C;
In the object measurement mode:

--Temperature readings too close together.

-- The object temperature is higher than 

100°C.

Hi: Higher than 100.0°C;

The 

thermometer 

show the 

symbol “Lo”

The hair and sweat prevent the temperature 

achievement.

Temperature hampered by an air 

fl

ux.

In the forehead measurement mode:

-- The measuring distance is too far.

-- Measured the other object, such as the air 

from the air conditioner.

Lo: Less than 32.0°C
In the object measurement mode:

-- The measuring distance is too far.

-- Have water vapor condenses on the lens.

Lo: Less than 0°C

The ambient temperature is beyond of range 

of measurement (10°C-40°C or 50°F-104°F)

Keep the thermometer in the room

Whose Temperature is (10°C-40°C or 

50°F-104°F) for 30 minutes

The sensor or hardware is damaged 

Excluding the possibility of 

temperature allowance 

fi

rst ,then send 

the device to your dealer for repair

Low batter, but you can still use it

Keep an eye on power and continue 

to use.

Lower battery, however you can’t use it

Replace the new battery.

14. SPECIFICATIONS

Device name

Infrared Thermometer

Model

Cherub Baby 4 in 1

Measurement mode

Forehead, ear and object temperature measurement modes

Power supply

d.c.3V, CR2032 batteries

Measuring range:

32.0-42.9°C (89.6-109.2°F)

Measuring range:

For human body temperature:

32.0°C-42.9°C (89.6°F-109.2°F)

For object temperature:

0°C-100°C (32°F-212°F)

Measuring accuracy:

(At laboratory conditions)

For human body temperature:

± 0.2°C/0.4°F  35. 5°C -42.0°C(95.9°F-107.6°F)

± 0.3°C/0.5°F  for other ranges

For object temperature: ±1.0°C/ 1.8°F  within 15-60°C(59.0°F-140.0°F);  

for other range , ±2.0°C/ 3.6°F

Clinical repeatability:

Within ±0.3°C

Resolution of display

0.1°C/0.1°F

Operation condition

15.0°C-40.0°C(59.0°F-104°F) Relative humidity: 20% to 95% RH

Transportation and storage 

condition

-25-55°C(-13°F-131.0°F), Relative humidity 

95%RH, 

Size

133.8

×

32.7

×

23mm

Weight

39g

High body temperature hint

38.0°C(100.4°F)

Grade of waterproof

IP22

Electric shock

Internally powered ME equipment

Applied part

Type BF applied part, including the whole unit

Mode of operation

Continuous operation

Note: Not intended to be sterilized.

Not for use in an OXYGEN RICH ENVIRONMENT

*The above speci

fi

cations are subject to change without prior notice.

Note: ASTM laboratory accuracy requirements in the display range of 36 to 39°C (96.8 to 102.2°F)

for this thermometer is ±0.2°C(0.4°F), whereas for mercury-in-glass thermometer, the 

requirement per ASTM standards E 667-86 is ±0.1°C(0.2°F).

15. STANDARD LIST

Famidoc Technology Co., Ltd declares that the FDIR-V1 complies with following applicable standards:

EN ISO 15223-1

Medical device – Symbols to be used with medical device labels, labeling and 

information to be supplied – Part 1: General requirements

EN 1041

Information supplied by the manufacturer with medical devices

EN 60601-1

Medical electrical equipment Part 1: General requirements for basic safety 

and essential performance

EN 60601-1-2

Medical electrical equipment -- Part 1-2: General requirements for basic 

safety and essential performance - Collateral standard: Electromagnetic 

compatibility - Requirements and tests

Note: 

FDIR-V1 provides both Fahrenheit and Celsius readings in countries where the Fahrenheit units 

are adopted, otherwise it provides only the Celsius unit. It isn’t a bad one if your product can’t 

change °C or °F.

Battery installment and replacement

1.   When the quantity of electric charge is not su

ffi

cient, the icon of low battery will appear on the 

lower part of the screen. Though the battery still can be used, it would be perfect to replace it.(Refer 

to the Picture 9.1)

2.   When the quantity of electric charge of the battery is lower than 

the minimal value, “Lo” symbols will be displayed on the screen, and 

battery icon 

fl

ashes and the sound of “Beep” is sent out (Refer to the 

Picture 9.2).Unless you change a new battery, you can not use the 

thermometer to conduct measurement.

3.   Operation of changing battery.

4.   Take out the old battery: open the battery cap (Refer to the Picture 9.3). 

Put a new battery on: put on a CR2032 lithium battery with cathode 

downward into the battery seat. (Refer to the Picture 9.4)

Notice:

t

 Please observe the related national laws of disposing the abondoned battery and don’t   

litter to the garbage can.

t

Please take out the battery if the device is not used for long periods of time.

To protect the environment, dispose of empty battery at your retail store or at appropriate collec-

tion sites according to national or local regulations.

8.  CARE AND CLEANING

Probe tip (lens) is a most precise part in the thermometer. Please keep clean and entirety in order to 

assure the accuracy of 

fi

gure.

The probe tip and lens are the most delicate part of the thermometer. It has to be clean and intact to 

ensure accurate readings.

Please use the ways to clean the probe as follows:

t

  Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the 

alcohol has completely dried out.

t

  If the lens is damaged, contact the distributor.

Clean the unit body:

t

  Use a soft, dry cloth to clean the thermometer display and unit body.

t

  If very dirty, use a soft with alcohol to cleaning.

NOTES:

t

  Don not use abrasive cleaners. 

t

  Don’t use other non-recommended methods to perform disinfect.

t

  Non-waterproof, don’t use the abrasive cleaner to clean the product, don’t drop the thermometer in 

the water or the other liquid.

9. MAINTENANCE

1)  We do not authorize any institution or individual to maintain and repair of the product. If you 

suspect that the products have any questions, please contact the manufacturer or distributor to 

handle the case.

2)  The user must not attempt any repairs to the device or any of its accessories. Please contact the 

retailer for repair.

3)  Opening of the equipment by unauthorized agencies is not allowed and will terminate any claim to 

warranty.

10. CALIBRATION

The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according 

to the use instruction, periodic calibration is not required. If any time your question the accuracy of 

measurement, please contact distributor or manufacturer, the contact information see last page.

11.  STORAGE

1)  Don’t put the thermometer under the sunshine, high temperature and moist environment or 

someplace which maybe get in touch with 

fi

re or is vulnerable to vibration.

2)  Take out the battery if don’t use the device in a long time.

12. ACCESSORIES

Only use original accessories. Check that the contents of the delivery are complete.

Quantity

Parts

1pc

Cherub Baby Thermometer

1pc

CR2032 battery(install in the thermometer)

1pc

User Manual

1pc

Base stand

13.  TROUBLE-SHOOTING

Troubles or 

error message

Checklists or situation

Countermeasures or solution

No response/ 

Automatically 

reset

The batteries are used up?

Replace new batteries.

Battery in wrong polarity or type?

Take out the batteries and replace new 

ones.

Take out batteries and reinsert it 

correct.

Poor battery contact

Edition:

 V2..1

Cherub Baby

9/6a Railway ave, Oakleigh VIC 3166

mail@cherubbaby.com.au

+61390051734

Product Manufactured by:

Famidoc Technology Co., Ltd.

Add.: No. 212 Yilong Road, Hexi Industrial Zone, Jinxia, Changan

Town, Donguang 523853, Guangdong Province, P.R. China.

Reviews: