Chapter 4
Configuring Settings
(
MODE SELECT Mode
)
46
4.3
Communication Format Setting Mode(When
Equipped with Communication Function)
This mode sets the below listed items of serial communication format.
The settings are enabled after the main power switch is turned off and back on.
Item
Setting
Default
Communication station number
1 to 25
1
Baud rate
9600 bps, 19200 bps,
38400 bps, 57600 bps,
115200 bps
9600 bps
Bit length and parity
Bit length = 8 bits, with no parity
Bit length = 8 bits, odd parity
Bit length = 8 bits, even parity
Bit length = 9 bits, with no parity
Bit length = 8 bits, with no parity
Stop bit
Stop bit = 1 bit
Stop bit = 2 bits
Stop bit = 1 bit
1.
Move to the MODE SELECT mode.
2.
Press the Up/Down arrow buttons to move to the communication format setting mode
(“Com Setting”).
3.
Press the ENTER button.
The communication station number check screen appears, showing the current settings.
4.
Press the Up/Down arrow buttons to move to the item you want to set.
5.
Press the ENTER button.
The setting screen appears.
6.
Press the Up/Down arrow buttons to select the desired setting.
7.
Press the ENTER button to determine the setting.
8.
To exit this mode, press the MODE SELECT button to return to normal mode.
4.4
Air Volume-Down Alert Timing Setting Mode
This mode allows for changing the timing for displaying an air volume-down warning (WARN4) as desired.
1.
Move to the MODE SELECT mode.
2.
Press the Up/Down arrow buttons to move to the air volume-down alert timing setting mode
(“Volume Down ST”).
3.
Press the ENTER button.
The current setting appears.
The default is 50%, displayed as “3: Down to 50%”.
4.
Press the Up/Down arrow buttons to select a timing between 30 to 70%.
A lower setting causes the air volume-down alert to occur later, and vice versa.
5.
Press the ENTER button to determine the setting.
6.
To exit this mode, press the MODE SELECT button to return to normal mode.
Summary of Contents for CBA-750AT3-HC-DSA-V1
Page 6: ...1 4 1 4 2 OFF 1 5...
Page 7: ...1 CMN034 002 Translation of the original instructions 5 1 6...
Page 12: ...2 10 2 4 3 OP kPa SP kPa DP kPa EP kPa Blower Runtime h 17 Total h Moter...
Page 13: ...3 CMN034 002 Translation of the original instructions 11 3 3 1 3 1 1 0 40 80 RH REMOVE 1 1 1...
Page 14: ...3 12 3 1 2 10 1 2 3 2 1 ON RS485 ID 2 AT3 ON 3 4 1 7 3 ON OFF 30 3 ON OFF 1 1 Ver ID...
Page 20: ...5 18 5 5 1 WARN2 1 5 1 1 1 1 1 2 2 1 3 1 4 2 1...
Page 22: ...5 20 3 4 5 HEPA 6 5 2 1 Littell fuse 218 250V 10A 1 2 5 3 1 1 1 1 5 4...
Page 26: ...6 24 6 6 1 6 1 1...
Page 27: ...6 CMN034 002 Translation of the original instructions 25 1 5V 0 2V DC 1 2...
Page 31: ...7 CMN034 002 Translation of the original instructions 29 7 3 7 3 1 CBA 750AT3 HC DSA V1...
Page 32: ...30 12 7 2 7 2 28...
Page 63: ...MEMO...