background image

User Guide

7

Servicing and Maintenance

Based upon EN10535 and health and safety executive guidance Chiltern
Invadex recommend that hoists are fully serviced, including a load test, at
least every 12 months with an interim six monthly inspection to LOLER
requirements. Prices are available from Chiltern Invadex Service Offices. The
frequency of servicing should be increased in areas where the environment
or heavy usage may cause deterioration of the equipment.  Further details
can be obtained from any Chiltern Invadex Office, shown on the back of this
booklet.

Chiltern Invadex and its Authorised Dealers are able to carry out all
aspects of servicing and maintenance work.
Maintenance of this equipment should only be conducted by engineers
trained and certified by Chiltern Invadex.

Keep the service record (at the back of this booklet) up to date.

DO NOT USE IF:

1. There is any damage to the hoist.

2. There is fraying or stitch damage to the lifting tape,

or slings.

To ensure maximum cleanliness
Wipe with a dry cloth (track and hoist). For stubborn marks use a damp cloth
and a mild cleanser. Always read the manufacturers instructions prior to
applying the cleanser.
Slings should be washed in accordance with the instructions on the label,
sewn into the sling.

Parts

item

description

qty

1

3 Button Hand Control           300-00120

1

2

5 Button Hand Control           300-00121

1

3

7 Button Hand Control           300-00122

1

1

2

3

Summary of Contents for Wispa 200 Series

Page 1: ...Overhead Hoists user guide Wispa 200 300 Series Series 200 300...

Page 2: ...you experience any difficulties please consult the trouble shooting guide prior to contacting your Overhead Hoist supplier Read the safety section of this guide prior to use this is contained on page...

Page 3: ...spa 200L 200LT 200LTX 200kgs 31 Stone 440 lb Wispa 300 272kgs 42 Stone 598 lb Ensure when the hoist is in use that the surrounding area is clear and the person being lifted is clear of all objects Saf...

Page 4: ...ervice engineer trained by Chiltern Invadex Do not touch any of the moving parts in the hoist unit or in the track Contact Chiltern Invadex for assistance Take care when attaching or removing a sling...

Page 5: ...d Control Five Button Hand Control Lowering Traverse Lowering Traverse Lowering Traverse Traverse X Y Reset Lifting Traverse X Y Carry Bar 7 Button Hand Controls Hand Control Socket Lifting Tape Indic...

Page 6: ...function buttons as the five button hand control with the addition of the tracking for the X Y system The tracking X Y function is controlled by the blue buttons on the handset The buttons are colour...

Page 7: ...eplaced if necessary at each service and after use Frequent testing of this device in non emergency situations will quickly discharge the battery Battery Low Indicator The battery low indicator will b...

Page 8: ...ealers are able to carry out all aspects of servicing and maintenance work Maintenance of this equipment should only be conducted by engineers trained and certified by Chiltern Invadex Keep the servic...

Page 9: ...position continuous red flashing 4 Low main 12V Battery 3 beeps pause Return to charge position continuous red flashing 5 Low main 24V battery 4 beeps pause Return to charge position continuous red f...

Page 10: ...s cycle with average 75kg 11 3 stone patient weight This is adjustable and should be set by the installer to meet the requirements of the user Typically the carry bar would stop about 300mm from the f...

Page 11: ...wear and tear wilful or accidental damage misuse neglect or any other cause which is beyond the control of Chiltern Invadex 4 Modifications and repairs made to this product by unauthorised persons wi...

Page 12: ...ngineers training cert no Signature Date Load Tested To Kgs Comments Serviced By company Telephone no Engineers training cert no Signature Date Load Tested To Kgs Comments Serviced By company Telephon...

Page 13: ...ephone no Engineers training cert no Signature Date Load Tested To Kgs Comments Serviced By company Telephone no Engineers training cert no Signature Date Load Tested To Kgs Comments Serviced By compa...

Page 14: ...Dealers Name Address Post Code Telephone Your Serial Number is Date of purchase Models covered 200 300 series Please fill in the information below for future reference...

Page 15: ...iltern Invadex Ltd Unit 16 Evans Business Centre 68 74 Queen ElizabethAvenue Hillington Industrial Estate Glasgow G52 4NQ Tel 0141 880 9986 Fax 0141 880 9990 Copyright Chiltern Invadex Ltd The design...

Reviews: