- 92 -
8-27. System Settings
“System”
The system related settings such as enabling/disabling settings are available. You can disable a setting change by keys
(Key Lock), clear memory (Initialize), disable/enable zero or span adjustment for dot printing position (Adjust of Rec
position) or disable/enable input adjustment (Input Correction) by entering a password. Set these items as a recovery
process when the unit does not function properly due to misoperation or other reasons.
(1) Pressing the
key displays the
menu window (list of setting items).
(2) Select
“System”.
(3) Pressing the
key opens the
password entry window.
(4) Enter a password. After that, move the
cursor to
and press the
key.
(5) When password entry is completed, move
the cursor to
and press the
key.
(6) When the setting window is displayed,
move the cursor to the parameter to be set
(7) Press the
key and check
the
desired item.
(8) After completing the settings of this item,
move the cursor to Set .
(9) Press the
key to register the
settings (when chart recording is ON, a
setting change mark is printed). To cancel
the settings, press the
key.
Set
Set
The default password is
“3571”. This password
cannot be changed.
Default password
Note 1
When the memory clear (Initialize) is executed,
parameters will be reset to default values.
This cannot initialize the adjustment data
(zero/span calibration).
Memory clear
Note 2
When Key Lock is enabled, settings of all items cannot be changed.
The message
“Key locking …..” is displayed when you try to register a parameter.
However, checking of setting items is available.
Key operation
Note 3