INST.No.INE-475-P1CE Software Version 1.00
-18-
6.4.1-1) Descriptions of setting data
Setting data name
Description
(1) Time constant
If the measured value fluctuates noticeably, the signal can be set to
be dull by delaying the response of this detector unit. The
smoothing time (equivalent to the time constant of analog
instruments) can be set in 0.1-second increment.
(2) Hold ON/OFF
By setting it to ON, the measured value is held.
At the same time, the analog output is also held.
(3)
Preset output
ON/OFF
By turning on the preset output, this detector unit displays the preset
value and the analog output becomes the value corresponding to the
preset value.
(4) Constituents
number
For the setting the number of constituents displayed. It skips in the
case of constituents 1.
The setting of constituent number 1 to 4 is possible and
displays it in
circulation with 0. Defalut value is 1.
(5)
Number of
decimal places
For the setting the number of decimal places of displayed data
(6)
Constituents number/CH
number display selection
Select constituent number/CH number display.
Set it to ON for
changing the CH display from
constituent number to calibration curve
number in measuring mode.
Defalut value is OFF.(constituent number
display)
(7) Constituents
number
of
analog output/Contact
output
The setting of constituent number of analog output and contact
output is possible. It skips in the case of constituents 1. The setting of
constituent number 1 to 4 is possible. Defalut value is 1.
(8) Contact output : Selection
of contact output
enabled/disabled and
output item
Select whether the contact output is enabled or disabled, and select
the output item.
(1) NonE
:Disabled
(2) ALArM :Outputs the high/low alarm, by the contact, of the
calibration curve number being set
(3) Error
:It is outputted when an abnormality is detected by the
self-diagnosis function shown in [9.2 Self-diagnosis
function].
(9) Contact input : Selection
of contact input
enabled/disabled and
output item
Select whether the contact input is enabled or disabled, and select the
output item.
(1) NonE: Disabled
(2) HoLd: The analog signal is held by the latest value.
(Holding by the signal H)
(3) PrSEt: The analog signal is set to a fixed value of 4-20mA.
(Fixing by the signal H)
(4) SMt.rL: Switching of Smoothing or Real (Real by signal H)
(10) Communication protocol rtU: MODBUS RTU mode
Priv: Exclusive mode - *This is not used with this detector unit.
(11) Detector
number
This is a number for identifying each detector in the
multiple-detector-units connection. This setting has been completed
in [6.2 Setting detector unit numbers].