INST.No.INE-475-P1CE Software Version 1.00
-24-
6.7.2 Procedure for setting data for output limit values of calibration curves
6.7.2-1 Setting the calibration curve numbers (CH)
1) Press SEL key and CH/. key simultaneously for 2 seconds to
display "CH" in the CH display for entering the calibration
curve number setting mode for the calibration curve output
limit values.
2) Set a calibration curve number.
ENT
6.7.2-2 Enabling/disabling the output limit processing (LM)
1) Press SEL key to enter the output limit processing
enabled/disabled "LM" setting mode.
Press key to enter a setting data.
Change the setting data by key, key, or key.
Press ENT key to store the setting data.
6.7.2-3 Setting a low limit absorbance value (XL) and a low limit measured value (YL)
1) After the above setting, press SEL key to enter the low limit
absorbance value "XL" setting mode.
Press key to enter a setting data.
Change the setting data by key, key, or key.
Press ENT key to store the setting data.
2) After the above setting, press SEL key to enter the low limit
measured value "YL" setting mode.
Press key to enter a setting data.
Change the setting data by key, key, or key.
Press ENT key to store the setting data.
6.7.2-4 Setting a high limit absorbance value (XH) and a high limit measured value (YH)
1) After the above setting, press SEL key to enter the high limit
absorbance value "XH" setting mode.
Press key to enter a setting data.
Change the setting data by key, key, or key.
Press ENT key to store the setting data.
2) After the above setting, press SEL key to enter the high limit
measured value "YH" setting mode.
Press key to enter a setting data.
Change the setting data by key, key, or key.
Press ENT key to store the setting data.
6.7.2-5 Setting output limit values to other calibration curve numbers
1) After the above setting (6.6.2-1 to 6.6.2-4), press SEL key to return to the calibration curve
number setting mode.
Enter other calibration curve number for setting output limit values.
Repeat the above procedure (6.6.2-1 to 6.6.2-4) to set output limit values to other calibration
curve numbers.
CH
DATA
X H 1. 0 0 0
CH
DATA
YL 0.0
CH
DATA
Y H 5 0. 0
CH
DATA
L M O F F
Setting data
Setting data name
CH
DATA
C H 1
Calibration curve number
setting for calibration
curve output limit value
Calibration
curve number
CH
DATA
XL 0.0
6. Operation
After the above setting procedure is completed, return to the measurement mode
by pressing SEL key for two seconds. The automatic return is made if no key is
pressed for one minute.
R e m a r k s