INST.No.INE-475-P1CE Software Version 1.00
-30-
8.1 General
When a calibration curve changes depending on the sample temperature Ts, this function enables to
switch the calibration curve to other one corresponding to the sample temperature measured. For the
input of the sample temperature, a radiation thermometer or a temperature/voltage converter with
the output of 4 to 20mADC can be used. By utilizing this function, 10 calibration curves in one
group (9 calibration curves only in the group 10) can be switched corresponding to the sample
temperature Ts. 99 calibration curves are available and up to 10 groups can be set accordingly.
8.2 Wiring this detector unit
8. Sample temperature correction function
Moisture content “Y”
Absorbance "X"
Calibration curve of sample
temperature "T3"
Calibration curve of sample
temperature "T2"
Calibration curve of sample
temperature "T1"
Correction input
4 to 20mA DC
P+
P-
OUT+ OUT-
SA
SB
SG
DI
COM
DO
COM
IN+
IN-
24VDC
4-20mADC
RS-485
DIGITALIN
DIGITALOUT
4-20mADC