INST No INE -403-0P0
−
33
−
8.6 Setting of sample temperature scaling (Only 4 to 20mA DC)
In case of using radio thermometer or thermoelectric converter, it is necessary to match the
temperature range of their 4 to 20mA DC output and the temperature range of 4~20mA DC
input of this instrument.
Carry out scaling in either detector unit side or setting display unit side, following the main
points given below.
8.6.1 Setting at the detector unit
(1) Enter into the operating condition setting mode.
SEL + ENT for 2 seconds
(2) Enter into the setting mode of sample temperature scaling low limit (tL).
SEL Several times
(3) Set the low limit (tL) of sample temperature scaling.
(Setting range: -999.9 to 999.9)
(4) Enter into the setting mode of sample temperature scaling high limit (tH). SEL Once
(5) Set the high limit (tH) of sample temperature scaling.
(Setting range: -999.9 to 999.9)
8.6.2 Setting at the setting display unit
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Enter into the sample temperature scaling setting mode
MODE 2 1 ENT
(3) Set the low limit (tL) of sample temperature scaling.
(Setting range: -999.9 to 999.9)
(4) Enter into the setting mode of sample temperature scaling high limit (tH). SEL Once
(5) Set the high limit (tH) of sample temperature scaling.
(Setting range: -999.9 to 999.9)
8.7 Setting of sample temperature
The key operation for this setting is common on the detector unit and the setting display unit.
(1) Enter into the calibration curve setting mode.
SEL for 2seconds
(2) Enter into the sample temperature setting mode (tS).
SEL several times
(3)Set the sample temperature.
(
Setting range: -999.9 to 999.9
)
8.
Correction function by sample temperature from external input