INST.No.INE-863B KR2S
【
Communication Interface
】
Instruction Manual
-47-
Reference
number
Applied
function code
R/W
Content
Detail
30186
04
R
CH43 decimal point/status
Same as CH1
30187
04
R
CH44 measured data
Same as CH1
30188
04
R
CH44 decimal point/status
Same as CH1
*About the exponent part form data format.
MSB(15) (14) (5) (4) (3) LSB(0)
Integer
part
sign
Integer part(10 bit)
Exponent
part
sign
Exponent part(4 bit)
Integer part sign
:sign
0(
+
) / 1(
-
)
Exponent part sign
:sign
0(
+
) / 1(
-
)
*About the error
1023
:Over
High limit over flow
1022
:Under
Low limit over flow
1021
:RJ
RJ error
1020
:BURN
Burnout
1019
:Cal
Calculation error
1018
:Other error
1017
:Invalid data
* About the decimal point/status information
MSB(15) (11) (7) (4) (3) LSB(0)
0
0
0
0
EV4
EV3
EV2
EV1
ERR
BURN
OF
UF
DP
EV1 to 4 : Each alarm status
0 (Not activated)/1 (activated)
ERR : Input part status
0 (normal)/1 (abnormal)
BURN : Sensor disconnection
0 (Not activated)/1 (activated)
OF : Overflow error
0 (Not activated)/1 (activated)
UF : Underflow error
0 (Not activated)/1 (activated)
DP : Data decimal place
|0|0|0|0|: 0, |0|0|0|1|: 1, |0|0|1|0|: 2, |0|0|1|1|: 3
Summary of Contents for KR2S
Page 2: ...INST No INE 863B KR2S Communication Interface Instruction Manual ...
Page 124: ...INST No INE 863B KR2S Communication Interface Instruction Manual ...
Page 125: ...INST No INE 863B KR2S Communication Interface Instruction Manual ...
Page 126: ...INST No INE 863B KR2S Communication Interface Instruction Manual ...
Page 127: ...INST No INE 863B KR2S Communication Interface Instruction Manual ...