Kinetec Kinevia™ / Kinetec Kinevia Duo™
EN
9/44
During cleaning and care:
◦ Clean the device between each patient, especially the parts in contact with patients such as the arm trainer grips (9)
and the support and/or transport bar (2).
◦ Do not use cleaning products containing solvents, acids or abrasives.
◦ It cannot be excluded that carpeting, wood flooring or tiling cleaning products contain substances that are corrosive
to the Kinevia support stand (7) plastic coatings. This could lead to deposits on the floor. Kinetec is not liable for this
damage.
◦ Regularly check that the screws are properly tightened.
During the entire device service life:
◦ Only the manufacturer or staff authorised by Kinetec are authorised to alter the device.
◦ Disconnect the power supply before any work on the device.
◦ Extensions, re-installations, alterations or repairs should only be carried out by staff authorised and approved by
Kinetec.
◦ If unexpected operation or events, strange noises, a smell of burning occur, immediately power off the device by
disconnecting the power cord and contact your Kinetec® dealer.
◦ All serious incidents related to the system must be the subject of a notification to the manufacturer and the
competent authority in the member state in which the user and/or patient are located.
◦ Avoid allowing liquids to penetrate inside the device or the touch screen.
◦ Kinevia is designed for use in the seated position. Do not climb onto the foot fixings with your whole weight and do
not hang from the arm trainer.
◦ The leg trainer must not be loaded with more than 25 kg on each side.
◦ The arm trainer must not be loaded with more than 15 kg on each side.
◦ If there is electromagnetic or other interference with other devices, more the system away from them.
◦ Kinevia must not be used next to other devices or stacked with them as this could lead to malfunctions. If this use is
required, observe this and the other devices to check that they are still in working order.
◦ The use of accessories and cables other than those indicated or supplied by Kinetec can cause increased
electromagnetic emissions or reduce device immunity and cause malfunctions.
◦ Portable radio-frequency communication devices (including peripherals such as aerial cables and external aerials)
should not be used closer than 30 cm (12 inches) from any part of the Kinevia, including the cables specified by
Kinetec. If this is not the case, the performances of those devices could be altered.