26/44
EN
Kinetec Kinevia™ / Kinetec Kinevia Duo™
Patient entry and exit aid
Entry aid
Before starting the training, you can use the entry/exit aid to move the foot supports to a comfortable and easy to
access entry position.
a)
Press the “Entry/Exit aid” key and keep it pressed until one of the foot supports is in an
easy to access position for the foot. A green light is displayed in the upper right corner of
the key.
b)
Release the “entry/exit aid” key. A red light is displayed in the upper right corner of the key.
The motor brake is activated.
c)
Place the foot in the foot support and attach it using the hook and loop strap.
d)
Press the “Entry/Exit aid” key again and keep it pressed until the second foot support is in
an easy to access position for the foot. A green light is displayed in the upper right corner
of the key.
e)
Release the “entry/exit aid” key. A red light is displayed in the upper right corner of the key.
The motor brake is activated.
f)
Place the second foot in the foot support and attach it using the hook and loop strap.
To change the direction of rotation, press the
key.
To disable the motor brake, press the
key.
Exit aid
To remove your feet once training is complete, repeat the same operations as for entry aid.
WARNING
RISK OF INJURY WHEN GETTING IN POSITION
Always place your feet in front of the device forward support when they
are not fixed in the foot supports.
Make sure your feet or legs will not be touched by the foot supports during
rotation.