38/44
EN
Kinetec Kinevia™ / Kinetec Kinevia Duo™
Cleaning
DANGER
DANGER OF DEATH BY ELECTROCUTION
Before cleaning the device, it is essential to switch off the device and
unplug the power cord.
Make sure that no cleaning or disinfection liquids penetrate inside the
device.
If electric components or cables are damaged, contact your Kinetec
dealer.
WARNING
HEALTH RISK FROM A CONTAMINATED DEVICE
Wear suitable protective gloves to clean and disinfect the device.
◦ Clean the device between each patient.
◦ Before each cleaning or disinfection, switch off the device and unplug the power supply.
◦ Clean the surface of the Kinevia by wiping using a soft and damp cloth.
◦ Disinfect the surfaces of the Kinevia (especially those in contact with the patient) using a surface disinfectant such as
Bactinyl
®
from Laboratoire Garcin-bactinyl
®
(Kinetec order references: 4665005483 for a spray or 4665005491 for
wipes).
◦ Do not spray cleaning or disinfection products directly on the device or its accessories.
◦ Make sure that no cleaning or disinfection liquids penetrate inside the device.
◦ When cleaning the touch screen, only use specifically adapted cleaning products.
◦ During cleaning and disinfection, make sure no stickers or protective films are damaged.
◦ Only use cleaning and disinfection products compliant with regulations and specific standards in the country of use.
◦ Only use cleaning and disinfection products that are not aggressive and are harmless to the environment.
◦ Let the disinfectant solution dry sufficiently before returning the device to service.
WARNING
RISK OF DAMAGE TO THE DEVICE THROUGH
INCORRECT CLEANING OR DISINFECTION
Do not use cleaning products that contain solvents, acids or abrasives as
they could attack the plastic parts of the screen.
Follow the safety instructions provided by the manufacturers of the
cleaning and disinfection products.
When in doubt, contact the cleaning or disinfection product manufacturer.
Maintenance and safety inspections
Kinetec recommends carrying out a visual inspection of the device before each training session.
Kinetec recommends subjecting the device to annual maintenance in order to check the condition of the belt as well as
different device elements. Please contact your Kinetec
®
dealer to carry out this operation.