Kinetec Kinevia™ / Kinetec Kinevia Duo™
●
Electromagnetic immunity, portable radio-frequency equipment
The
device
is
designed
for
use
in
the
magnetic
and
electromagnetic
environment described in the table below.
Users and installers must guarantee the compliance of the electromagnetic environment.
WARNING
MALFUNCTION RISK
Portable radio-frequency communications devices (including peripherals
such as aerial cables and external aerials) should not be used closer than
30 cm (12 inches) from any part of the device, including the cables
specified by Kinetec. If this is not the case, the performances of those
devices could be altered.
Immunity tests
Test level
Compliance level
Electromagnetic environment - Remarks
Radiated radio-
frequency
electromagnetic fields
IEC 61000-4-3
3 V/m
from 80 MHz to 2.7 GHz
80% MA at 1kHz
10 V/m
from 80 MHz to 2.7 GHz
80% MA at 1kHz
3 V/m
from 80 MHz to 2.7 GHz
80% MA at 1kHz
10 V/m
from 80 MHz to 2.7 GHz
80% MA at 1kHz
The device is suitable for use in a home
healthcare environment and in a professional
healthcare establishment environment.
Proximity fields
emitted by wireless
radio-frequency
communications
devices
IEC 61000-4-3
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
9 V/m
710 MHz, 745 MHz,
780 MHZ, 5240 MHz,
5550 MHz, 5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz, 810 MHz,
870 MHz, 930 MHz,
1720 MHz, 1845 MHz,
1970 MHz, 2450 MHz
The device is suitable for use in a home
healthcare environment and in a professional
healthcare establishment environment.
Conducted
interference induced
by radio-frequency
fields
IEC 61000-4-6
3 V
from 150 kHz to 80 MHz
6 V in the ISM band and
band included between
0.15 MHz and 80 MHz,
including the amateur
radio band
80% MA at 1 kHz
3 V
from 150 kHz to 80 MHz
6 V in the ISM band and
band included between
0.15 MHz and 80 MHz,
including the amateur
radio band
80% MA at 1 kHz
The device is suitable for use in a home
healthcare environment and in a professional
healthcare establishment environment.
The electromagnetic field intensity of static radio-frequency emitters as determined by an electromagnetic environment
measurement (a), must be less than the conformity level for each frequency range.
Interference can occur near equipment identified by the following symbol.
REMARK: These specifications may not apply in all situations. Electromagnetic propagation is influenced by the absorption
and reflection of structures, objects and persons.
(a) The electromagnetic field intensity of static radio-frequency emitters such as mobile phone base stations (cell / cordless),
mobile radios, amateur radios, AF/FM radio emissions and TV emissions cannot be accurately determined by the theory.
To assess the electromagnetic environment due to the static radio-frequency emitters, an electromagnetic environment
measurement must be carried out. If the measured intensity of the radio-frequency field in the immediate product use
environment exceeds the radio-frequency conformity level specified above, the product performances must be tested to
check that they are compliant with specifications. If abnormal performances are found, additional measurements may be
needed, such as reorienting or moving the product.