EN

 

Kinetec Kinevia™ / Kinetec Kinevia Duo™ 

Electromagnetic immunity, portable radio-frequency equipment

The 

device 

is 

designed 

for 

use 

in 

the 

magnetic 

and 

electromagnetic 

environment described in the table below. 
Users and installers must guarantee the compliance of the electromagnetic environment. 

WARNING 

MALFUNCTION RISK 

Portable  radio-frequency  communications  devices  (including  peripherals 
such as aerial cables and external aerials) should not be used closer than 
30  cm  (12  inches)  from  any  part  of  the  device,  including  the  cables 
specified  by  Kinetec.  If  this  is  not  the  case,  the  performances  of  those 
devices could be altered. 

Immunity tests 

Test level 

Compliance level 

Electromagnetic environment - Remarks 

Radiated radio-
frequency 
electromagnetic fields 
IEC 61000-4-3 

3 V/m 

from 80 MHz to 2.7 GHz 

80% MA at 1kHz 

10 V/m 

from 80 MHz to 2.7 GHz 

80% MA at 1kHz 

3 V/m 

from 80 MHz to 2.7 GHz 

80% MA at 1kHz 

10 V/m 

from 80 MHz to 2.7 GHz 

80% MA at 1kHz 

The  device  is  suitable  for  use  in  a  home 
healthcare  environment  and  in  a  professional 
healthcare establishment environment. 

Proximity fields 
emitted by wireless 
radio-frequency 
communications 
devices 
IEC 61000-4-3 

9 V/m 

710 MHz, 745 MHz, 

780 MHZ, 5240 MHz, 

5550 MHz, 5785 MHz 

27 V/m 

385 MHz 

28 V/m 

450 MHz, 810 MHz, 
870 MHz, 930 MHz, 

1720 MHz, 1845 MHz, 

1970 MHz, 2450 MHz 

9 V/m 

710 MHz, 745 MHz, 

780 MHZ, 5240 MHz, 

5550 MHz, 5785 MHz 

27 V/m 

385 MHz 

28 V/m 

450 MHz, 810 MHz, 
870 MHz, 930 MHz, 

1720 MHz, 1845 MHz, 

1970 MHz, 2450 MHz 

The  device  is  suitable  for  use  in  a  home 
healthcare  environment  and  in  a  professional 
healthcare establishment environment. 

Conducted 
interference induced 
by radio-frequency 
fields 
IEC 61000-4-6 

3 V 

from 150 kHz to 80 MHz 

6 V in the ISM band and 

band included between 
0.15 MHz and 80 MHz, 

including the amateur 

radio band 

80% MA at 1 kHz 

3 V 

from 150 kHz to 80 MHz 

6 V in the ISM band and 

band included between 
0.15 MHz and 80 MHz, 

including the amateur 

radio band 

80% MA at 1 kHz 

The  device  is  suitable  for  use  in  a  home 
healthcare  environment  and  in  a  professional 
healthcare establishment environment. 

The  electromagnetic  field  intensity  of  static  radio-frequency  emitters  as  determined  by  an  electromagnetic  environment 
measurement (a), must be less than the conformity level for each frequency range. 

Interference can occur near equipment identified by the following symbol.

 

REMARK:  These specifications  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  influenced  by  the  absorption 
and reflection of structures, objects and persons. 
(a) The electromagnetic field intensity of static radio-frequency emitters such as mobile phone base stations (cell / cordless),

mobile radios, amateur radios, AF/FM radio emissions and TV emissions cannot be accurately determined by the theory.
To  assess  the  electromagnetic  environment  due  to  the  static  radio-frequency  emitters,  an  electromagnetic  environment
measurement  must  be  carried  out.  If  the  measured  intensity  of  the  radio-frequency  field  in  the  immediate  product  use
environment exceeds the radio-frequency conformity level specified above, the product performances must be tested to
check that they are compliant with specifications.  If abnormal performances are found, additional measurements may be
needed, such as reorienting or moving the product.

Summary of Contents for Kinetec Kinevia

Page 1: ...inals auf Franz sisch Im Falle einer Abweichung wird die Franz sisch Original ma gebend IT Istruzioni per l uso Prima di mettere in funzione l apparecchio leggere con attenzione il presente documento...

Page 2: ......

Page 3: ...E KINETEC KINEVIA DUO UPPER AND LOWER LIMB TRAINER 11 DETAIL VIEW OF THE TOUCH SCREEN 12 ACCESSORIES 12 Calf support 12 Self handling foot attachment 12 Tilt protection hooks 12 Heart rate kit 13 Ergo...

Page 4: ...NING POSSIBLE USING KINEVIA 29 QUICK START 29 GUIDED TRAINING THERAPY PROGRAMS 29 Training phases 31 Training levels 31 INDIVIDUAL TRAINING 32 TRAINING ANALYSIS 32 TRAINING STATISTICS 34 SETTINGS 34 T...

Page 5: ...ication Symbol used to warn of hazards independently of the danger level The danger level is indicated by the corresponding warning word as described in section 1 1 Prohibition Symbol used to indicate...

Page 6: ...infections Brittle bones Unstable circulation Side effects The therapy should be discussed with the attending physician the therapist or the kinesitherapist concerning its extent and duration in order...

Page 7: ...ion of an adult Kinevia is designed for humans with a body weight of up to 135 kg The arm trainer must not be loaded with more than 15 kg on each side The leg trainer must not be loaded with more than...

Page 8: ...nce which favours muscular development Resistance can be set to 20 different levels comparable to changing gears on a bicycle The higher the level the harder it is to pedal This function is used to de...

Page 9: ...e sure that the power supply block and the electric cables are in good condition Only use the original power cord supplied with the device Do not use accessories spare parts or supplies other than tho...

Page 10: ...ice height setting the pedalling radius the distance to the pedals and all training parameters are set to match your physical capacities Make sure that the legs or arms are properly fixed before start...

Page 11: ...ystem must be the subject of a notification to the manufacturer and the competent authority in the member state in which the user and or patient are located Avoid allowing liquids to penetrate inside...

Page 12: ...If assistance is needed for safe training make sure the assistant is in the immediate vicinity during training so that they can stop the Kinevia at all times WARNING DANGER DUE TO PHYSICAL DEFICIENCIE...

Page 13: ...transport bar Grip height indexing button On Off switch Foot pedal Power supply plug Support stands Transport castors General view of the Kinetec Kinevia Duo upper and lower limb trainer The following...

Page 14: ...leg movements Order reference 4665009477 Self handling foot attachment The self handling foot attachments are used to hold the feet in place in the supports Locking and unlocking are easy to access th...

Page 15: ...515 Support glove The fixture for a Handfix wrist is used to attach the hands to the arm trainer grips It is recommended in cases of spasms hand paralysis and all hand motor pathologies which prevent...

Page 16: ...sport and settings Unpacking and installation Please follow the safety instructions on page 7 Before starting up remove all packaging and transport protection Check that the device is complete and ide...

Page 17: ...mly hold the device using the transport grips with both hands f If rolling the Kinevia to the required spot is not possible for example due to thresholds or stairs two people should carry it by liftin...

Page 18: ...us can be continuously set between 50 and 125 mm a Please power off the device before setting the pedal radius b Slightly unscrew the wing nuts located at the outside end of the pedal arms photo c Gui...

Page 19: ...UNTIGHTENED INDEXING BUTTON Before each training session check that the indexing button is anchored and tightened Make sure the grips do not collide with the legs when setting their height Calf suppo...

Page 20: ...n the 2 pivoting arm locking screws 11 d You can now activate the arm trainer from the touch screen see section 8 3 2 on page 23 Kinevia Duo use as a leg trainer Kinevia Duo use as an arm trainer It i...

Page 21: ...t sunlight for long periods Standby mode Activating standby mode Standby mode is activated by pressing anywhere on the screen for about 5 seconds Exiting standby mode Standby mode is disabled by press...

Page 22: ...raining for the first time ask your doctor or therapist show you the correct seated position General training advice Training using Kinevia has positive effects on the body and mind Regular training c...

Page 23: ...page 29 Guided training accesses the pre programmed training selection screen for various pathologies see section 9 2 on page 29 Individual training accesses the selection of other training without pr...

Page 24: ...ck to your last setting or to the previous screen Validate key This key is used to confirm a menu selection Cancel key This key is used to cancel a menu selection Start key Pressing this key starts th...

Page 25: ...m up training recovery Home key Used to go back to the home page Lower command panel displays and keys Automatic change of rotation direction key This key is used to enable or disable the automatic ch...

Page 26: ...Leg trainer mode selection key This key is used to select the leg trainer or arm trainer operating modes The key is enabled if and only if the arm trainer is properly installed and leg training has b...

Page 27: ...ee section 8 6 3 on page 29 This key is disabled for passive training Symmetry This field respectively indicates the load level for the left and right sides of your body during active training The ind...

Page 28: ...Entry Exit aid key again and keep it pressed until the second foot support is in an easy to access position for the foot A green light is displayed in the upper right corner of the key e Release the...

Page 29: ...ss the ASC key to access the settings screen accessible when the motor is stopped or running b Press the plus minus keys to set the spasm detection sensitivity 0 spasm detection disabled 1 weak spasm...

Page 30: ...duration to the physical capacities and state of health of the person who is training Short but regular training sessions should be preferred Rotation speed Pressing the rpm key accesses the rotation...

Page 31: ...ing possible using Kinevia Quick start Using this option you directly access the training menu see section 8 3 on page 23 and you can compose your own therapy programme by selecting the passive soft a...

Page 32: ...s uses a target heart rate The motor resistance is automatically reduced once the heart rate has been reached You can adapt the target heart rate by pressing the heart rate key DANGER DANGER OF DEATH...

Page 33: ...training training entirely using the motor drive force Rotation speed can be changed at all times Recovery at the end of the training phase a short recovery phase for passive training and at a reduce...

Page 34: ...eak resistance to 20 very high resistance depending on your personal needs Starting training using a low resistance is recommended Training analysis At the end of the training session you can view you...

Page 35: ...or resistance value in Newton metres Muscle tone Nm max This field indicates the maximum motor resistance value in Newton metres Resistance This field indicates the average motor braking resistance du...

Page 36: ...ues displayed as total for the analysis duration Values displayed as average values for the analysis duration Values displayed as maximum values for the analysis duration All the training values are a...

Page 37: ...me must be reset if the Kinevia has been powered off User data This menu is used to set patient data using the arrows to move from field to field and the plus minus keys to change values such as age h...

Page 38: ...tion pedal rotation resumes in the opposite direction Forward pedal rotation resumes in the forward direction Back pedal rotation resumes in the reverse direction Stop pedal rotation stops Change of d...

Page 39: ...annot be repaired or if the repair costs exceed the current value of the device compensation limited to the current value of the device will be paid Exclusions from the warranty extension the warranty...

Page 40: ...ids penetrate inside the device When cleaning the touch screen only use specifically adapted cleaning products During cleaning and disinfection make sure no stickers or protective films are damaged On...

Page 41: ...s per minute for active training Distance between foot rests from 13 5 to 25 5 cm measured on the inner sides Distance between arm grips 12 cm measured on the inner sides Pedal radius Leg trainer Arm...

Page 42: ...vice double insulation Device reference Do not sit on the device BF TYPE device protection from electric shocks Device serial number Keep hands away moving parts Alternating current Manufacture batch...

Page 43: ...resistance during pedalling The braking resistance in active mode is too high Reduce the braking resistance Activate soft training if necessary 8 6 3 Spasm detection triggers too quickly or too often...

Page 44: ...RISK If there is mutual electromagnetic or other interference with other devices move the device away from them The device must not be used next to other devices or stacked with them as this could le...

Page 45: ...st electric transients bursts IEC 61000 4 4 2 kV for electricity supply lines 2 kV for electricity supply lines The quality of the electricity supply grid must be equivalent to that of a home healthca...

Page 46: ...in a professional healthcare establishment environment Conducted interference induced by radio frequency fields IEC 61000 4 6 3 V from 150 kHz to 80 MHz 6 V in the ISM band and band included between 0...

Page 47: ......

Page 48: ......

Reviews: