Kinetec Kinevia™ / Kinetec Kinevia Duo™
EN
41/44
Troubleshooting guide
Defect
Possible cause
Action
See
section
The Kinevia will not start
The power cord is not plugged in
Plug the power cord into a mains
power outlet and to the power
supply box.
Plug the power supply box cable
into the Kinevia connector.
The device is not switched on
Turn the on/off switch to “I”.
Standby mode has been
activated
Press anywhere on the screen
for about 5 seconds.
Defective fuse
Contact the maintenance
technician
The movements are not fluid
Pedal radius setting different on
each side
Set the pedal radius to the same
value on both sides
Incorrect seated position
Correct the seated position
Too high resistance during
pedalling
The braking resistance in active
mode is too high
Reduce the braking resistance.
Activate soft training if necessary
Spasm detection triggers too
quickly or too often
The spasm detection level setting
is too sensitive
Set the spasm detection (ASC)
level to a higher value
The arm trainer will not start
Arm trainer not activated on the
training menu
Press the arm trainer selection
key
Leg training not complete
Stop leg training by pressing the
Stop key before selecting the arm
trainer
The heart rate display is not
working
The Kinevia cardio kit is not
installed or is not properly
installed
Connect the receiver properly to
the jack socket. The logo should
point towards the front.
Check and correct the thoracic
belt position.
If you need to return the device or its accessories, first discuss the return and packaging specifications with Kinetec.