English
19
Set-Up and Starting
The following describes the setting up and the starting of the SONIC vital
XTcoloR.
1.
The device should be placed on a smooth, even surface.
2.
Please ensure that the SONIC vital XTcoloR is not placed immediately adjacent
to other devices and neither should it be stacked on other devices. This is to
avoid reciprocal interference. If this is not possible, the function of the respec-
tive devices must be kept under observation to ensure that proper operation
is not compromised.
3.
Voltage supply connection
The connection for the mains power cable is located on the reverse side of the
device. It must be checked that the mains voltage supply values are in compli-
ance with the values stated on the rating plate. Once this has been checked,
the mains cable is to be connected to the device and plugged into the mains
socket.
4.
Connecting the ultrasound heads
Connect the required ultrasound head(s) to any of the connectors on the
reverse side of the device. When removing the ultrasound head from its
socket never pull on the cable: always grasp the plug casing! Otherwise, the
ultrasound head could be damaged or ruined.
5.
The device is switched on with the ON/OFF switch which is located on the
reverse side of the device.
6.
The start video screen will appear after a few seconds.
7.
Once the main menu appears in the display, the device is ready to operate.