English
29
Technical Data
Protection Class
II
Application Parts
Type BF 2 pieces
Operations Type
Continuous operation
Dimensions
43 cm x 24 cm x 10 cm
Weight
1.7 kg (only the device)
Voltage Supply
100 - 240 VAC
Power Take-Up
0.18 A
Duty Cycle (Tact Ratio)
10 - 50 % in 5 Steps (Step Width 10 %), 100 %
Maximum Power
2 W/cm² (duty cycle = 100 % = continuous
operations) and 3 W/cm² in duty cycle operations
Pulse Frequency
100 Hz.
Number of Programs
60 Fixed and 10 User Programs
Treatment Duration
0 - 59 Min.
Ultrasound Head
SVDF 1 (small size)
Acoustic Working Frequency 1 and 3 MHz
Power Density
2 W/cm² and 3 W/cm² in duty cycle operations
ERA
1 cm²
BNR
< 4.0
Bundle Type
Collimating
Ultrasound Head
SVDF 4 (large size)
Acoustic Working Frequency 1 and 3 MHz
Power Density
2 W/cm² and 3 W/cm² in duty cycle operations
ERA
4 cm²
BNR
< 4.0
Bundle Type
Collimating
Classification
The SONIC vital XTcoloR has been classified as a Class IIa device in compliance with
the classification criteria of the Medical Device Directive 93/42/EU, Appendix IX,
Rule 9.