c) Mount the receiver boxes in the ceiling and install the disinfecting modules in the receiver
boxes (on page 27).
8. Turn on power at the source.
When the disinfecting modules are inserted in the receiver box and power is on at the source,
the modules are automatically powered on and start flashing red indicating the modules are
not connected to a network.
9. In the CounterAct application:
a) Log into the CounterAct application (on page 20).
b) Create a room (on page 40).
c) Pair multiple devices in the application (on page 40).
When pairing devices after installation, ensure all the MAC addresses are recorded in
advance to distinguish the devices.
d) Add the devices to the room (on page 41).
e) Configure the scheduler (on page 41) to determine when you want disinfection to occur.
10. When all the devices are installed, configured, and turned on, run the devices in Disinfection
mode and using a meter, verify the optical power output (on page 41) are below the ACGIH
Risk Group Exempt levels for the space.
When Disinfection mode is not on, the device is in Standby mode.
Planning the layout and power of the site
Planning the site can help determine the installation position and power to help facilitate pairing and
configuration work.
Christie strongly recommends planning out the site before arriving at the site for installation.
1. Make a plan of the installation position for each device.
For example, the site plan should show the planned location of all the devices.
2. Make a plan for the power to be used at the site.
Planning the installation height
Plan the installation height for CounterAct.
The mode lamp mode the device will be run in depends on the installation height:
•
Installed in a ceiling with a minimum height of 6.7 m (22 feet) to 8.5 m (28 feet)—Single-lamp
mode
•
Installed in a ceiling with a minimum height of 8.5 m (28 feet)—Dual-lamp mode
Below 6.7 m (22 feet) this model of the device should not be used. Contact Christie Sales for options
suitable for installation in ceilings below 6.7 m (22 feet).
Installation and setup
CounterAct Installation and Setup Guide
18
020-103456-01 Rev. 1 (09-2021)