12. Carefully route the wires and maintain the gap between the wires and the wires to the metal
cover.
The wires (A in the image below) must sit in center of grommets (B in the image below) and
maintain a minimum distance of 8 mm (0.3 inches) between the wires and the wires to the
metal shield.
13. Connect the lamp connector to the integrator board.
14. Re-install the disinfecting module (on page 14).
15. Reset the lamp hours.
a) Ensure you are logged into CounterAct Configure with the installer account.
b) From the homepage, select Fixture.
c) From the Fixture page, select Reset life.
d) From the Reset life dialog, select the lamp you want to reset.
The fixture life is reset.
16. Ensure you dispose of lamps according to local regulations.
Maintenance and service
CounterAct Installation and Setup Guide
46
020-103456-01 Rev. 1 (09-2021)