Manual Operation
3-59
Figure 3-102
2.
When the address of to be calibrated SLAVE is confirmed, keep in parallel mode and go
to the CONFIG setting page, press
“
7
” and “
ENTER
” to enter into CALIBRATION
for password input as shown in Figure 3-100.
3.
Select MASTER OR SLAVE via
“Rotary Knob” (
) and select the SLAVE to be
calibrated as Figure 3-103 shows, press
“
ENTER
” to enter into the calibration screen as
Figure 3-104 shows. As to the calibration procedure of each item, please see the
descriptions from section 3.3.7.1 to 3.3.7.3.
Figure 3-103
Figure 3-104
1.
Password is required for CALIBRATION. The password is “3636”.
2.
The SLAVE model can only do calibration via MASTER. First
remove the output connecting cable and then disconnect the current
sharing cable except ANALOG and SYSTEM BUS cables which
should be kept on the device.
3.
The MASTER is for SLAVE panel output only, thus the load and
measurement cable for actual calibration needs to be installed on
SLAVE.
4.
The DC Power Supply should be calibrated regularly and it is