Manual Operation
3-91
Figure 3-163
6.
Press
“
ON/OFF
” again for execution. It will jump to the SAS MODE main screen as
shows. To quit, press “
EXIT
” to return to the MAIN PAGE.
Figure 3-164
Figure 3-164 shows the screen when the SAS MODE outputs. The EQ_PMP, EQ_VMP
and EQ_IMP are the actual operating points that are calculated from the settings of user
VMP and IMP via the hardware built-in formula.
The user can change the VOC, ISC, VMP and IMP parameters directly using the “
”,
“
” function keys from the screen of single unit. When any of the VOC, ISC, VMP or
IMP is modified,
use “
” to move the cursor to TRIG and press “
ENTER
” to confirm.
The IV curve will update immediately and so too the EQ_PMP, EQ_VMP and EQ_IMP
parameters.