Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
2-2
1.
Please keep all of the packing materials in case the device has to
be returned for repair.
2.
Do not return the instrument to the factory without obtaining prior
RMA acceptance from Chroma.
3.
Check if the accessories listed in the packing list are all received.
CAUTION
The power supply is too heavy for one person to safely lift and mount. To
avoid injury, ask a co-worker for assistance.
2.1.1
Maintenance & Cleaning
Remove all connected wires and cables from the instrument before cleaning. Use a brush
to clean the dust on it and if there are stains on the chassis that cannot be removed by brush,
wipe it with volatile liquid (such as Cleaning Naphtha). Do not use any corrosive liquid to
avoid damaging the chassis. Use a damp cloth with soap water or soft detergent to clean the
LCD front panel. For internal cleaning, use a low-pressure air gun for the dust inside or
send it back to our agent for cleaning.
2.2
Preparation for Use
(1) Remove the iron holder of the front panel as Figure 2-2 shows and keep it in case the
Power Supply is required for return service.
(2) Be sure the Power Supply is connected to the AC line input that meets the specification.
(3) The instrument must be installed in an air-circulated area to avoid the internal
temperature getting too high.
(4) The ambient temperature does not exceed 40
C.
Figure 2-2
CAUTION
1.
If the iron holder on the front panel is not removed, it may cause
OTP or damage on the hardware due to poor ventilation.
2.
The 62020H-150S model has no such device.