Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-34
[O U T P U T S E T U P]
V L I M I T : M A X = 6 0 0 . 0 _ V M I N = 0 . 0 V
Σ
I L I M I T : M A X = 2 5 . 0 0 A M I N = 0 . 0 A
V S L E W R A T E = 1 . 0 0 0 ( V / m S )
Σ
I S L E W R A T E = I N F . ( A / m S )
V D C _ R = 6 . 0 V V D C _ F = 6 . 0 V
▼
Figure 3-54
3.3.3.5.4
Setting PROTECTION for Parallel
The OCP and OPP in PROTECTION of MASTER for parallel connection will increase
following the number connected in parallel. It is indicated by
ΣOCP and ΣOPP for easy
identification as Figure 3-55 shows below.
[ P R O T E C T I O N ]
O V P
= 6 0 0 . 0 _ V
∑
O C P
= 1 8 . 9 0 A
∑
O P P
= 1 0 5 0 0 . 0 W
R E M O T E I N H I B I T = D I S A B L E P U L L = H I G H
S A F E
T Y I N T . L O C K = D I S A B L E P U L L = H I G H
E X T O N / O F F
= D I S A B L E P U L L = H I G H
Figure 3-55
3.3.3.6
Setting Procedure for APG in Series or Parallel
3.3.3.6.1
Series Setting
To connect 2 sets of 62150H-600S DC Power Supplies in series for operation and set the
APG option to APG VSET/APG ISET = Vref(0-5V), the MAIN PAGE of MASTER will show as
Figure 3-56.
∑ V = 1 2 0 0 . 0 0 _ V I = 2 5. 0 0 A M S T O F F
0 . 0 0 0 0 V 0 . 0 0 0 0 A
0 . 0 W
A P G
–
V S A P G
–
I S
Figure 3-56