Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-78
Table 1-1 for selecting the appropriate BREAKER.
3.3.7.7.2
SETUP
The connecting diagram for current calibration is shown in Figure 3-141.
62020H- 150 S
v
v
DVM
sv
sv
Electronic
Load
Breaker
ULTRASTAB
SATURN
Figure 3-141
1.
Before using ULTRASTAB SATURN for current calibration, it needs
to calibrate the offset voltage of ULTRASTAB SATURN first. Press
“
Null
” and the DVM panel digits will reset and show “Math” for
current calibration.
2.
When conducting the current calibration, each calibration point has
to key in at least 5 Arabic numerals to ensure the Power Supply
accuracy after calibration.
3.3.7.7.3
Calibration Procedure
1.
Enter into the screen as shown in Figure 3-136.
2.
In the CALIBRATION page, press
“
2
” or turn the “Rotary” (
) knob to set
CHOICE=2.
3.
Press
“
ENTER
” to confirm entering into the current calibration as Figure 3-142 shown
below.
[ C U R R E N T M E A S C A L I B R A T I O N ]
R E M O V E A L L L O A D I N G F R O M O U T P U T T E R M I N A L
P R E S S [ E N T E R ] W H E N R E A D Y
Figure 3-142