Manual Operation
3-87
2.
Use the numeric
0
~
2
keys
or “Rotary” (
) knob to select the OUTPUT
MODE. The default is CV/CC MODE.
3.
Press “
ENTER
” to confirm.
4.
Press “
EXIT
” to return to the MAIN PAGE.
3.3.9.1
CV/CC MODE
In addition to the function of Solar Array Simulation the models of Chroma 62000H Series
DC Power Supplies with Solar Array Simulation can also be used as general DC Power
Supply. Thus the operating mode still has CV mode and CC mode as Figure 3-157 shows.
See section 3.2 for detail operating procedures.
CV1
CV2
CC1
CC2
Figure 3-157
3.3.9.2
TABLE MODE
To use this function in a DC Power Supply it has to work with the Solar Array Simulation Soft
Panel. The user can edit the IV Curve data in Excel and use the software to import the IV
Curve data to DC Power Supply for Solar Array Simulation. The mode allows the user to
edit various IV Curves as Figure 3-158 shows to generate the Solar Cell IV Curves of many
types or the IV Curves under different operating environments such as irradiation change,
temperature change, cloud shadow etc. It can also change the curve in real time to
facilitate the user when using the IV Curves under various conditions.
(a) (b)