Program Sequence
4-1
4.
Program Sequence
62000H Series DC Power Supply with Solar Array Simulation allows the user to program the
sequence for output in LIST MODE, V_STEP MODE and IV PROGRAM. LIST MODE and
IV PROGRAM both have 10 Programs and each Program can add new sequences liberally
that total 100 sequences, while IV PROGRAM is used to edit the IV Curve waveform in
particular. V_STEP MODE provides a run time voltage program with the maximum of 99
hours.
Each sequence in LIST MODE can be edited for voltage settings, voltage slew rates, current
settings, current slew rates, run times and trigger types that can apply to almost any
situation.
1.
Press “
PROG
” on the front panel.
2. It displays PROGRAM options as Figure 4-1 shows.
3. Use numeric keys
1
~
3
or “Rotary” (
) knob to set the desired mode.
4.
Press “
ENTER
” to confirm.
5. To quit PROGRAM, just press
“
EXIT
” to return to the MAIN PAGE.
P R O G R A M M O D E = L I S T _ [ P R O G R A M ]
1. L I S T M O D E
2. V _ S T E P M O D E
3. I V P R O G R A M
Figure 4-1
4.1
LIST MODE
In LIST MODE there are a maximum of 100 sequences that can be added liberally in one
program. The sequence setting is described in section 4.1.2 and the complete program
structure is listed in Figure 4-2.