Program Sequence
4-17
[ P R O G R A M / S T E P]
S T A R T _ V O L T A G E = 0 . 0 0 V
E N D _ V O L T A G E
= 0 . 0 0 V
R U N _ T I M E
= 0 : 0 : 0 . 0 0
▼
Figure 4-18
4.2.1
Setting V_STEP MODE
V_STEP MODE has 3 settings: (1) START_VOLTAGE, (2) END_VOLTAGE and (3)
RUN_TIME.
4.2.1.1
Setting START_VOLTAGE
1. Use
“
”, “
” keys to move the cursor to the column to be set as Figure 4-19
(1) shows. Set the start voltage of STEP MODE.
2. Use numeric keys
0
~
9
or “Rotary” (
) knob to set the value.
3.
Press “
ENTER
” to confirm.
4.
Press “
EXIT
” to return to Figure 4-1.
[ P R O G R A M / S T E P]
S T A R T _ V O L T A G E = 0 . 0 0 _ V
E N D _ V O L T A G E
= 0 . 0 0 V
R U N _ T I M E
= 0 : 0 : 0 . 0 0
▼
( 1 )
( 2 )
( 3 )
Figure 4-19
CAUTION
The initial voltage of the hardware does not equal to the setting of the
START_VOLTAGE. There are two circumstances that may occur in
V_STEP MODE: (1) The output voltage rises to the setting of
START_VOLTAGE and the V SLEW RATE is 1V/mS, or (2) it falls to the
setting of START_VOLTAGE and the falling time is calculated by 1V/mS
while the actual V SLEW RATE is varied by load.