Manual Operation
3-7
0V ~ 600V/0A ~ 25A will map to 0~10V as Figure 3-7 (b) shows.
c. When selecting Iref=4-20mA
it means the DC Power Supply’s
output 0V ~ 600V/0A ~ 25A will map to 4-20mA as Figure 3-7 (c)
shows.
d. When selecting Vref=5KOhm
it means the DC Power Supply’s
output 0V ~ 600V/0A ~ 25A will map to 0~5KOhm as Figure 3-7 (d)
shows.
D
C
S
O
U
R
C
E
S
E
T
600V/25A
APG INPUT 5V
D
C
S
O
U
R
C
E
S
E
T
APG INPUT 10V
600V/25A
(a) (b)
D
C
S
O
U
R
C
E
S
E
T
600V/25A
APG INPUT 4 ~ 20mA
D
C
S
O
U
R
C
E
S
E
T
APG INPUT 5KOhm
600V/25A
(c) (d)
Figure 3-7
2.
APG VMEAS/APG IMEAS
has 4 selections which are NONE /
Vref(0-5V) / Vref(0-10V) / Iref(4-20mA) where:
a. When selecting Vref=5V
it means the DC Power Supply’s
measurement output 0 ~ 600V/0A ~ 25A will map to 0~5V as Figure
3-8 (a) shows.
b. When selecting Vref=10V
it means the DC Power Supply’s
measurement output 0 ~ 600V/0A ~ 25A will map to 0~10V as Figure
3-8 (b) shows.
c. When selecting Iref=4~20mA
it means the DC Power Supply’s
measurement output 0 ~ 600V/0A ~ 25A will map to 4mA~20mA as
Figure 3-8 (c) shows.