Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-32
3.3.3.4.3
Setting OUTPUT SETUP for Series
The V LIMIT MAX in OUTPUT SETUP of MASTER for series connection will increase
following the number connected in series. It is indicated by
ΣV LIMIT MAX: for easy
identification as Figure 3-50 shows below. Furthermore, the setting range of
ΣV SLEW
RATE will increase following the number connected in series too.
[O U T P U T S E T U P]
Σ
V L I M I T : M A X = 6 0 0 . 0 _ V M I N = 0 . 0 V
I L I M I T : M A X = 2 5 . 0 0 A M I N = 0 . 0 A
Σ
V S L E W R A T E = 1 . 0 0 0 ( V / m S )
I S L E W R A T E = I N F . ( A / m S )
V D C _ R = 6 . 0 V V D C _ F = 6 . 0 V
▼
Figure 3-50
3.3.3.4.4
Setting PROTECTION for Series
The OVP and OPP in PROTECTION of MASTER for series connection will increase
following the number connected in series. It is indicated by
ΣOVP and ΣOPP for easy
identification as Figure 3-51 shows below.
[ P R O T E C T I O N ]
∑
O V P
= 6 0 0 . 0 _ V
O C P
= 1 8 . 9 0 A
∑
O P P
= 1 0 5 0 0 . 0 W
R E M O T E I N H I B I T = D I S A B L E P U L L = H I G H
S A F E
T Y I N T . L O C K = D I S A B L E P U L L = H I G H
E X T O N / O F F
= D I S A B L E P U L L = H I G H
Figure 3-51
3.3.3.5
Setting Parallel Parameters
When the software communication and hardware settings for parallel are completed, the
settings of following windows are introduced in the sections underneath - (1) MAIN PAGE, (2)
SYSTEM SETUP and (3) OUTPUT SETUP.
3.3.3.5.1
Setting MAIN PAGE
MAIN PAGE is mainly used to set voltage (V) and current (I). The difference between single
unit and parallel operation is that the current set will increase following the number
connected in parallel. The current set is indicated by
ΣI for easy identification. When set