Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-42
This function sets the protection point for Over Current. Once the output current exceeds
the range, it will turn off the output that is OUTPUT = OFF to protect the unit under test.
Table 3-3 shows the current range of OCP.
Table 3-3
Model
Min. OCP (A)
Max. OCP (A)
62xxxH-xxxxS
0
1.05 x Io_MAX
When OCP occurs the main page will prompt a protection message as Figure 3-72 shows
below:
V = 6 0. 0 0 V I = 1 0. 0 0 _ A O F F
0 . 0 0 0 0 V 0 . 0 0 0 0 A
O C P 0 . 0 W
Figure 3-72
3.3.5.3
OPP Protection
1.
Use “
”, “
” keys to move the cursor to the column to be set as Figure
3-73 shows.
[ P R O T E C T I O N ]
O V P
= 6 0 0 . 0 _ V
O C P
= 1 8 . 9 0 A
O P P
= 5 2 5 0 . 0 W
R E M O T E I N H I B I T
= D I S A B L E P U L L = H I G H
S A F E
T Y I N T . L O C K
= D I S A B L E P U L L = H I G H
E X T O N / O F F = D I S A B L E P U L L = H I G H
▼
Figure 3-73
2.
Use the numeric (
0
~
9
) keys or “Rotary” (
) knob to set the value.
3.
Press “
ENTER
” to confirm.
4.
Press “
EXIT
” to return to the MAIN PAGE.