Manual Operation
3-55
3.3.5.15 C/S CABLE ERR. Protection
This protection occurs under the following three conditions. The first condition is when the
DC Power Supply is in single unit mode and the rear panel connects the CURR. SHARING
cable as Figure 3-95 shows. The second condition is when the DC Power Supply is in
series mode and the CURR. SHARING cable is connected as shown in Figure 3-96 and
Figure 3-97. The third condition is when the DC Power Supply is in parallel mode but the
CURR. SHARING cable is not connected as shown in Figure 3-96 and Figure 3-97.
Figure 3-95
Figure 3-96
Figure 3-97
Troubleshooting:
(1) When C/S CABLE ERR. protection occurs, please turn off the
Power Supply first, remove the load and check if the connections
are correct and then power it on again.
(2) This protection only supports when the hardware is upgraded and
the firmware version is 2.00 or above also the FPGA is 1.11 or
above. Please check if there is HOOP mounting hole on the rear
panel (the SLAVE models A620028 and A620027 are not included)
as Figure 2-10 shows on the hardware to make sure it is an
upgraded version.