Programmable DC Electronic Load 63600 Series Operation & Programming Manual
7-14
7.3.5.3 Checking Dynamic Constant Current Low Range
A. First, press
till VFD shows
CC
and press
to light up the L range LED
indicator.
B. Press
to enter into Dynamic Constant Current setting. The settings are listed in
C. Set the DC Source output voltage to 5V and the limit current needs to be larger than the
highest level set by dynamic constant current. (EX: When the 63630-600-15 H level is
set to 0.15A, the DC Source limit current can set to 0.5A.)
Table 7-19
Model
H
L
T1
T2
SR/
SR\
63630-600-15
150mA
0A
10ms
10ms
15m A/
μs 15m A/μs
63640-150-60
0.6A
0A
10ms
10ms
0.06 A/
μs 0.06 A/μs
63630-80-60
0.6A
0A
10ms
10ms
0.06 A/
μs 0.06 A/μs
63610-80-20
0.2A
0A
10ms
10ms
0.02 A/
μs 0.02 A/μs
63640-80-80
0.8A
0A
10ms
10ms
0.08 A/
μs 0.08 A/μs
Figure 7-3
MODE
RANGE
EDIT