Verification
10-3
10.4 AC Performance Test
10.4.1 I monitor & CF Measurement Verification
, please switch AC Load to CF only mode. The
procedures are described as follows.
1. Press
CONF./LOCAL
.
2. Press
1
and
ENTER
to enter SETUP. (More information about the display, please
refer 3.6.1 BOTH mode).
3. Move the cursors of
▲
and
▼
to “1-5. CF/PF SETTING”
.
4. Press direction keys of
◄
,
►
or turn rotary knob to select “CF” and then press
ENTER
.
(More information about the display, please refer 3.6.1 BOTH mode).
5. Press
CC
to return to main screen.
6. AC Source settings are shown as
, switch DVM measurement terminal to Rear.
Please follow the model under CC mode to set loading current, I
p
(max) & CF. According to
the model to confirm if the values of DVM, Meter & Load meet to that in
Table 10-2
Source
Setting
Load Setting
DVM
Meter & Load
Measurement
Model
Voltage
(V)
Frequency
(Hz)
Current
(A)
Ip(max)
CF
I_Mon
Specification
CF Specification
100
45/440
5
54
1.414
0.229~0.271
1.364~1.464
63802
5
4.949~5.051
18
3
0.876~0.925
2.95~3.05
5
108
1.414
0.229~0.271
1.364~1.464
63803
5
4.949~5.051
36
2
1.771~1.829
1.95~2.05
5
135
1.414
0.229~0.271
1.364~1.464
63804
5
4.949~5.051
45
1.415
2.219~2.281
1.365~1.465
Source setting in
, the frequencies of 45Hz and 440Hz are
tested once for each one. The settings in test item are the same.
10.4.2 CC Mode Setting & Measurement Verification
PF/CF BOTH mode settings are as follows.
1. Press
CONF./LOCAL
.
2. Press
1
and
ENTER
to enter SETUP
.