Local Operation
3-13
I s e t = 0 . 5 0 0 C F = 1 . 4 1 4 P F = 1 . 0 0 0
I p ( m a x ) = 1 0 . 0
U N I T : A
V = 0 . 0 0 P = 0 . 0 0
I = 0 . 0 0 S = 0 . 0 0
A C P R I O R I T Y : C F
If user chooses Lock ALL thus all keypads and rotary are disabled except for
LOCK
key.
3.5
AC Load Simulation
The 63800 series electronic load has AC/DC simulated functions. The AC load simulation
includes constant load mode and rectified load mode and its operation modes are described
as below.
In AC mode, the
CC
,
CR
,
CP
,
RLC/DC RECT
keys are used to select the load modes for
local control. AC mode and DC mode can be chosen in the 63800 series. Press
CONF./LOCAL
and select the first item, “SETUP”, to choose one. In AC mode, the load
values, PF and CF are common to CC, and CP modes; CR and RLC mode only set
determinate values. It is necessary to set UUT’s output maximum current value I
p
(max) in all
modes. The value I
p
(max) is for 63800 to test UUT’s output impedance for getting system
optimum.
The UUT’s output meets AC mode and frequency is within 45Hz ~
440Hz, output voltage 45Vrms ~ 350Vrms and output inductance is less
than 2mH thus it can normal use.
63800’s power-on default value is AC mode.
If user wants to return to AC mode from DC mode.
1. Press
CONF./LOCAL
key.
2. Press
1
, then press
ENTER
to enter SETUP.