Local Operation
3-31
R s = 0 . 2 0 L s = 1 0 0 C = 2 3 0 0 R L = 5 0 0 0 . 0 0
I p ( m a x ) = 1 0 . 0 U N I T :
Ω
, u H , u F , A
7. Move the cursor by pressing (
◄
,
►
) to the command line of “R
L
=”.
R s = 0 . 2 0 L s = 1 0 0 C = 2 3 0 0 R L = 5 0 0 0 . 0 0
I p ( m a x ) = 1 0 . 0 U N I T :
Ω
, u H , u F , A
8. Press
4
,
5
,
0
,
0
then press
ENTER
to change the value to “4500”.
R s = 0 . 2 0 L s = 1 0 0 C = 2 3 0 0 R L = 4 5 0 0 . 0 0
I p ( m a x ) = 1 0 . 0 U N I T :
Ω
, u H , u F , A
9. Move the cursor by pressing (
◄
,
►
) to the command line of “I
p
(max)=”.
R s = 0 . 2 0 L s = 1 0 0 C = 2 3 0 0 R L = 4 5 0 0 . 0 0
I p ( m a x ) = 1 0 . 0 U N I T :
Ω
, u H , u F , A
10. Press
3
,
0
then press
ENTER
to change the value to “30.0”.
R s = 0 . 2 0 L s = 1 0 0 C = 2 3 0 0 R L = 4 5 0 0 . 0 0
I p ( m a x ) = 3 0 . 0 U N I T :
Ω
, u H , u F , A
11. After setting the parameter of the load, the
LOAD ON/OFF
key should be pressed to turn
on the load, so that the load can sink the current. (For more detailed information please
refer
section 3.4.5
If there is error during the procedure of inputting value, press
CLEAR
to clear
the value at any time and then input again.
3.9.2 CP Mode
The appropriate parameter values of R
s
, L
s
, C, R
L
can be calculated by the algorithm of P
set
and PF settings. The value will be used to execute loading and the loading power will be
compensated automatically to be within the error range. The UUT I
p
(max) needs to be set
before loading.
For switching to CP mode, the procedures are as below:
1. Press
RLC/DC RECT
key to go into the Rectified Load Mode.
2. Press
2
, then press
ENTER
to enter CP mode.