Local Operation
3-43
I s e t = 2 5 . 0 0 0 C F = 1 . 4 1 4
I p ( m a x ) = 5 0 . 0 T I M E O U T = 0 0 : 0 1 : 0 0
V = 0 . 0 0
C U T O F F V = 5 0 . 0 0 0
T
= 0 0 : 0 0 : 0 0 : 0 0 0 m s
H O L D U P
A C
C F O N L Y
For switching to PF only mode, the procedures are as follows.
1. Press
CONF./LOCAL
key.
2. Press
1
, then press
ENTER
to enter SETUP.
3. Move the cursor (
▲
,
▼
) to “1-5. CF/PF SETTING”.
4. Press the direction key (
◄
,
►
) or rotate the rotary knob to choose “PF”, then press
ENTER
.
5. Then press
CC
to back to main page.
Users can press the arrow key to select the setting parameters on the LCD display, including
current RMS value, PF, I
p
(max), Timeout and Cutoff voltage. Users can type the values with
the Entry keys.
I s e t = 2 5 . 0 0 0 P F = 1 . 0 0 0
I p ( m a x ) = 5 0 . 0 T I M E O U T = 0 0 : 0 1 : 0 0
V = 0 . 0 0
C U T O F F V = 5 0 . 0 0 0
T
= 0 0 : 0 0 : 0 0 : 0 0 0 m s
H O L D U P
A C
P F O N L Y
3.10.2 CR Mode
Once under timing mode, if users want to switch to CR mode of timing mode press
CR
key
then to enter. There are three sub-modes under CC mode: BOTH, CF ONLY and PF ONLY.
CR mode
page is shown as below. Users can press the arrow key to select the setting
parameters on the LCD display, including resistor value, I
rms
(max), Timeout and Cutoff voltage.
Users can type the values with the Entry keys.