100673 REV
7
Intended Use
This SleepWeaver mask is intended to provide an
interface for Continuous Positive Airway Pressure
(CPAP) or Bi-Level therapy. This mask is intended for
single-patient reuse in the home and multi-patient,
multi-use in the hospital/institutional environment.
This mask is to be used on patients greater than
66 lbs (30 kg).
Cautions
• Refer to your therapy device manual for details on settings and operational
information.
• The technical specifications of the mask are provided for your clinician to check
that they are compatible with the therapy device. If used outside specification,
the seal and comfort of the mask may not be effective, optimum therapy may
not be achieved, and leak or variation in the leak rate may affect device function.
• The mask should be cleaned every other day to prevent dirt and oil
contamination from the skin building up on the mask.
• The mask should not be cleaned with unapproved cleaning or disinfecting methods.
• The user should clean their face prior to wearing the mask.
• The user should not alter the design of the mask.
• Do not place the mask in a clothes dryer.
• In the USA, federal law restricts this device to sale by, or on the order of, a
physician.
Contraindications
• None
Limited Warranty www.circadiance.com/help
White hook tab
attaches to the
top of mask
Feather
Weight
Tube
90 Degree
Elbow Swivel
Blue hook tab attaches
to the Bottom of mask
Exhalation
Holes
Size
Identification
Printing
Bottom of
Mask
Adjustment
Tab
1
2
3
4
5
Fitting the Mask
Mask Orientation
Technical Specifications
1.
Therapy Pressure Range: 4 to 20 cmH
2
O
2.
22mm Swivel Connection
3.
Resistance (pressure drop)
• @ 50 lpm = 0.4 cmH
2
O
• @ 100 lpm = 0.7 cmH
2
O
4.
Fixed leak rate (flow may vary due to manufacturing variation):
5.
Environmental storage and operating conditions are not expected to effect
the function of this device.
6.
Sound Levels
• A-weighted Sound Power Level,
27
dBA
• A-weighted Sound Pressure Level
@ 1m
,
19
dBA
7.
This device is expected to remain suitable for use under normal
usage and cleaning conditions for a time period of at least 90 days.
8.
The device is not made with natural rubber latex.
Pressure (cmH
2
0)
4
5
10
15
20
Flow (lpm)
14
16
27
37
45
English
Instructions for Use
Soft Cloth Masks
Élan
Mailing Address: PO Box 17480, Pittsburgh PA 15235
Physical Address: 1300 Rodi Road Turtle Creek PA 15145
Phone: 888-825-9640 or +1 724-858-2837
cservice@circadiance.com www.sleepweaver.com
www.circadiance.com/help
Watch the Fitting Video
®
EMERGO EUROPE
Prinsessegracht 20,
2514 AP The Hague,
The Netherlands
0459