Intended Use

This SleepWeaver mask is intended to provide an 

interface for Continuous Positive Airway Pressure 

(CPAP) or Bi-Level therapy. This mask is intended for 

single-patient reuse in the home and multi-patient, 

multi-use in the hospital/institutional environment.  

This mask is to be used on patients greater than  

66 lbs (30 kg).

Cautions

• Refer to your therapy device manual for details on settings and operational

information.

• The technical specifications of the mask are provided for your clinician to check

that they are compatible with the therapy device. If used outside specification,

the seal and comfort of the mask may not be effective, optimum therapy may

not be achieved, and leak or variation in the leak rate may affect device function.

• The mask should be cleaned every other day to prevent dirt and oil

contamination from the skin building up on the mask.

• The mask should not be cleaned with unapproved cleaning or disinfecting methods.

• The user should clean their face prior to wearing the mask.

• The user should not alter the design of the mask.

• Do not place the mask in a clothes dryer.

• In the USA, federal law restricts this device to sale by, or on the order of, a

physician.

Contraindications

• None

Limited Warranty   www.circadiance.com/help

White hook tab 

attaches to the 

top of mask 

Feather 

Weight 

Tube

90 Degree 

Elbow Swivel

Blue hook tab attaches 

to the Bottom of mask 

Exhalation 

Holes 

Size  

Identification 

Printing

Bottom of 

Mask

Adjustment  

Tab

1

2

3

4

5

Fitting the Mask

Mask Orientation

Technical Specifications

1.

Therapy Pressure Range: 4 to 20 cmH

2

O

2.

22mm Swivel Connection

3.

Resistance (pressure drop)

• @ 50 lpm = 0.4 cmH

2

O

• @ 100 lpm = 0.7 cmH

2

O

4.

Fixed leak rate (flow may vary due to manufacturing variation):

5.

Environmental storage and operating conditions are not expected to effect

the function of this device.

6.

Sound Levels

• A-weighted Sound Power Level, 

27

 dBA

• A-weighted Sound Pressure Level

 @ 1m

19

 dBA

7.

This device is expected to remain suitable for use under normal

usage and cleaning conditions for a time period of at least 90 days.

8.

The device is not made with natural rubber latex.

Pressure (cmH

2

0)

4

5

10

15

20

Flow (lpm)

14

16

27

37

45

English

Instructions for Use

Soft Cloth Masks

Élan

Mailing Address: PO Box 17480, Pittsburgh PA 15235
Physical Address: 1300 Rodi Road Turtle Creek PA 15145
Phone: 888-825-9640 or +1 724-858-2837
cservice@circadiance.com   www.sleepweaver.com

www.circadiance.com/help

Watch the Fitting Video

®

EMERGO EUROPE

 

Prinsessegracht 20,  

2514 AP The Hague,  

The Netherlands

0459

For patents and trademarks applicable to Circadiance products, please visit www.Circadiance.com/IP

Reviews: