– 11 –
b) 10 Amp Fast acting 6.3x32mm Fuse (by a professional person
only)
➔
Turn off the meter and disconnect both test leads.
➔
Remove the two screws at the back of the case.
➔
Remove the single screw securing the battery/fuse cover and lift to
open.
➔
Remove the batteries.
➔
Carefully lift the top of the case away from the bottom.
➔
The 10 amp fuse is the larger fuse on the bottom of the PC board,
which should now be in full view. Carefully remove the old fuse and
replace it with a new CSA/UL listed fast acting 10A/250V fuse only
(AGC-10A).
MAINTENANCE
a) Always keep the meter dry.
b) Keep the meter clean. Wipe the case occasionally with a damp cloth.
Do not use chemicals, cleaning solvents or detergents.
c) Use and store the meter in recommended normal environmental
conditions. Extreme temperatures can shorten the life of the electronic
components.
d) Use only fresh batteries.
e) Remove the batteries when the meter is not being used for a long
period of time.