28
4. High voltage indicator
An indicator (led)
is warning the presence of high voltage
at the output terminal during a measurement and remains lit
until the discharge process is completed. When you press
Stop
key, the equipment will start discharging the
potentials accumulated in the apparatus, internal
capacitances and in the element under test as well. When
this discharging process is over, the
high-voltage
led will
turn off automatically. The test leads may be disconnected.
5. Operating instructions
1. Be sure that there are no voltage differences between the points to
which the equipment will be connected to neither between them and
ground.
2. Connect the red test lead to the
5 kV
,
2.5 kV
,
1 kV
or
0.5 kV
terminal in accordance with the desired test voltage.
3. Connect the black test lead to the
-R
instrument terminal.
Connection without GUARD test lead.