CR:270 Series User Manual
Page 14
If this cannot be achieved the microphone or instrument may be at fault and a check
should be made with Cirrus Research.
As 'A' and 'C' weightings all have nominal zero correction loss at 1kHz, the reading on these two
positions should be nominally the same. However, due to minor instrument tolerances there will
be small divergences so it is recommended that calibration is performed on the actual weighting
which will be used for measurement.