background image

CR:270 Series User Manual 

 

 

Page 8 

In the right position, marked >80 the instrument measures levels in the range 80 to 

140dB.  
 

Thus, the total measuring range is about 32 to 140dB(A) for an instrument fitted with 
the MK:202 Microphone Capsule, and 27 to 140dB(A) for an instrument fitted with the 

MK:224 Microphone Capsule. Please note that the lower figure will depend on the 

microphone capsule in use. See section 10 for the instrument specifications. 
 

Response 

 
This switch defines the three response speeds "Slow",  "Fast" and "Impulse".  The IEC 

651 standard now calls these 'S', 'F' and 'I' respectively, so that they are the same in 

any language. 
 

On the "I" response it should be noted that the rise time constant is 35mS but the 
descent time constant is 1.5 seconds, thus it should only be used for single impulses. 

When in the "Impulse Max" position there is no decay time ie when the instrument is 

set to Impulse Time Weighting, and the Max Hold button is pressed. 
 

Display 

 
This is a three position switch.  In the centre position, coded PEAK(C) the CR:27x will 

measure C weighted peak Pressure, in the right hand position 'SPL' and in the left hand 

position it will measure Leq. 
 

The position of the switch does NOT affect the acquisition of data by the Leq  circuitry, 
the switch simply determines what is displayed on the meter. Thus the switch can be 

moved between the three positions while a measurement is being made. Indeed, it is 

quite normal to leave the switch in the SPL position to see how the noise is varying and 
then switch to Leq at the end to see what the average level was or to PEAK for the peak 

value since last reset. 
 

Max Hold 

 

When this switch is pushed in, the maximum SPL level as selected by the response 
switch is held and automatically updated by a further incoming signal.  This is 

conventionally called 'Max Hold'. The hold functions has a decay time of better than 1 
dB per minute.  

 

The Max Hold function can be reset whilst viewing by pressing the RESET DISPLAY 
button. 

 
The Max Hold function is automatically reset when the Display switch is moved from 

PEAK to Max Hold.  Therefore if the user wishes to make simultaneous measurements 

of Max Hold with Leq or Peak then the Max Hold reading should be recorded before the 
display switch is moved to Peak or Leq. 

 

Reset Display  

 

This button, when pressed, resets the function that has been selected for display.  

Summary of Contents for CR:270 Series

Page 1: ...CR 270 Series User Manual Page 1 CR 270 Series Sound Level Meters User Manual This manual the software to which it relates the program code and drawings are all Copyright Cirrus Research plc 1989 2015...

Page 2: ...plc No liability is accepted for any inaccuracies or omissions in this manual although due care has been taken to ensure that is it complete and accurate as possible Accessories supplied by Cirrus Re...

Page 3: ...0 Selecting the Active filter 10 Turning off Octave filters 11 Filters and Weighting 11 Display and Indicators 11 Annunciators 11 Full Centre colon 11 Overload 11 Battery low 11 Summary 12 Calibration...

Page 4: ...eting both these specifications to Type 1 and Type 2 accuracy where applicable The CR 273 and CR 274 instruments also features octave band filters that meet or exceed the requirements of IEC 1260 The...

Page 5: ...t Switches for Calibration Calibrate Instrument Check Battery Select Function Check Range Reset Sound Level SPL Average Sound Level Leq Peak Level LPeak C Record Measurement Check Calibration Power Of...

Page 6: ...sure that the black HOLD button is OUT If this is not done the display will not freeze the last reading and the central colon will be displayed in the LCD Always check the instrument calibration befor...

Page 7: ...ll be reset by the instrument to prevent nonsensical results due to mixtures of levels or weightings The centre position marked dB C permits use as a normal sound level meter with the C Weighting filt...

Page 8: ...he right hand position SPL and in the left hand position it will measure Leq The position of the switch does NOT affect the acquisition of data by the Leq circuitry the switch simply determines what i...

Page 9: ...ALL accidently In the Leq mode the button does NOT cause the reading to go to zero While RESET ALL is pressed the reading will be in an undefined random position as the meter is trying to divide two...

Page 10: ...e upper most label will all be extinguished Switching on the filters One quick push duration less than 1 second of the OCTAVE FILTER button will switch in the octave filters ie make them active Once t...

Page 11: ...is simply that data acquisition stops The CR 27x will still function on SPL and PEAK C If this colon lights then read the Leq note it along with the peak if required and reset the instrument Overload...

Page 12: ...rument more internationally compatible Summary Function Annunciator Cause Possible Action Overload Amplifier overload Move UP one range Under range Amplifier at noise floor Move DOWN one range Battery...

Page 13: ...psule whilst the CR 271 and CR 273 instrument are fitted with the MK 224 capsule Both of these microphones have a calibration correction of 0 3dB Refer to the Acoustic Calibrator manual for details Th...

Page 14: ...th Cirrus Research As A and C weightings all have nominal zero correction loss at 1kHz the reading on these two positions should be nominally the same However due to minor instrument tolerances there...

Page 15: ...ll the battery leads The batteries are connected to clips which are designed to prevent the batteries from being connected with the wrong polarity Types of batteries The CR 270 series should be used w...

Page 16: ...sses involved and the order in which these should be applied if appropriate Octave Band Method The Octave band method is a straightforward noise reduction calculation involving the measurement of the...

Page 17: ...s from the octave band sound pressure levels Octave Band Centre Frequency Hz 63 125 250 500 1000 2000 4000 8000 SPL APV 7 4 10 0 14 4 19 6 22 8 29 6 38 8 34 1 To calculate the effective A Weighted so...

Page 18: ...predicted noise level reduction PNR is calculated from the H M and L values The formula used depends upon the difference between the LCeq and the LAeq If LCeq LAeq is less than or equal to 2 2 4 A C...

Page 19: ...ncidence is parallel to the case as marked with an arrow on the cone IEC 804 651 11 2 3 The range of measurement for Leq and SPL measurement with a standard assembly is CR 272 CR 274 32 to 140 dB A 50...

Page 20: ...on for a microphone extension cable on the standard CR 272 CR 274 units However if an extension cable option is in use with the CR 271 or CR 273 units C irrus Research cables up to 10 meters long will...

Page 21: ...ld use IEC 804 11 2 28 The indicator range is identical to the linearity range of IEC 804 11 2 4 Additional Octave information The CR 274 meets IEC 1260 Type 2 whilst the CR 273 meets IEC 1260 Type 1...

Page 22: ...74 60 to 110dB 0 5dB CR 273 50 to 110dB 0 5dB IEC 1260 7j The range should be selected such that neither the underrange or overload flags occur during measurement IEC 1260 7k Not applicable IEC 1260 7...

Page 23: ...CD display battery annunciator indicates that the batteries are not sufficient to run the instrument correctly IEC 1260 7t The filter is integral to the CR 273 and CR 274 IEC 1260 7u The instrument is...

Page 24: ...ally 1 Gigohm Weighting networks A and C for CR 272 and CR 271 A and Lin for CR 274 CR 273 Display type 3 digit reflective LCD Time weighting S Slow F Fast I Impulse All to IEC 651 relevant Type Displ...

Page 25: ...after 1st January 1996 CR 272 Sound Level Meter CR 271 Sound Level Meter CR 274 Sound Level Meter CR 273 Sound Level Meter Along with their standard accessories According to EMC Directives 89 336 EEC...

Page 26: ...will be a reduced charge for their replacement during the annual Traceable Calibration 8 On completion of the annual Traceable Calibration by Cirrus Research plc or an official Cirrus Calibration Cent...

Page 27: ...manby North Yorkshire United Kingdom YO14 0PH Telephone 44 0 1723 891655 Fax 44 0 1723 891742 E mail sales cirrusresearch co uk Web Site www cirrusresearch co uk Germany Cirrus Research plc Deutschlan...

Reviews: