Indication of defected pixels (Address 104)
Operation when setting data and operation when acquiring data are different.
(1)
When setting (SU 104)
This is the function to indicate the actually corrected pixels by that time, with 75% brightness.
On/Off is done by parameters.
Parameter 0: OFF
Parameter 1: ON
Bright point indication function cannot be saved.
(2)
When acquiring (GU 104)
This is the function to indicate the registered defected pixels.
3 kinds of lists can be indicated by parameters.
Parameter 1: Entire lists, both ex-factory data, and the added and registered data by the users, are shown.
Parameter 2: Only the list added and registered by the users is shown.
(Example of indication)
From the left, X coordinate, Y coordinate, and the type of defects
[Type of defects]
W: White defects registered at Ex-factory
B: Black defects registered at Ex-factory
U: White defects registered by the user
P: Defects added by the user
Parameter 3:
Indicates the number of the registered defected pixels per CH (Data at Ex-f Data
registered by the user.)
Parameter 4:
Indicates the number of the registered defected pixels per CH (Data registered by the user
only.)
Parameter 5:
Error status of the registered data. Indicates if the data registered by the user are appropriate
One of the messages below shall be shown.
925 443 W
1228 460 W
1271 488 W
500 500 P
1098 520 W
930 629 U
OK: Normal
[ERROR] too many user defective pixels: Data number registered by the user exceeds the maximum (256 points).
[ERROR] region data full: Data SUM registered by the user and the one at ex-factory exceeds the limitation for 1 CH.
[ERROR] too many user defective pixels/region data full: Both of the above 2 cases occur at the same time.